Set Trigger Mode Command Kg

COMMAND <Kg mode_global, mode_port_1, mode_port_2, #>

Parameter

ASCII

Option

Mode

0

Continuous Read

Mode

1

Continuous Read, Send Once, Clear on Timeout

Mode

2

External Trigger, Read While Active

Mode

3

External Trigger, Read Once

Mode

4

Serial Trigger

Mode

5

Serial Trigger or External Trigger,

 

 

Read Once *

Sending the configuration command Kg sets the trigger mode. The command string is interpreted left to right. For example, if you send the following command: <Kg 4, 0>, both ports will be set to mode 4, then port 1 changes back to mode 0. Likewise, if you send the following command: <Kg , 0, 4>, port 1 is set to mode 0 and port 2 is set to mode 4. Both commands have the same end result; port 1 will be in continuous read mode and port 2 will be in serial trigger mode. The default setting is Mode_global 5, selecting a serial trigger or an external trigger, one read.

Tip 1: Careful consideration must be taken when configuring multiple ports to a continuous trigger mode. The 3700 decodes from only one port at a time. With multiple ports triggered, the 3700 will select each in turn and attempt to read. If no bar code label is present at the selected port, the 3700 will wait for read timeout to expire before stepping to the next port. If read timeout is set to a value near or greater than the time it takes a bar code label to pass by a scanner, then the label will often be missed.

Set Read Timeout Command Kh

COMMAND <Kh mode_global, time_global, mode_port_1, time_port_1, mode_port_2, time_port_2, #, #>

Parameter

ASCII

Option

Mode

0

Timeout Ends a Read Attempt * (150)

Time

 

one to five ASCII digits,

 

 

range of 10 to 65000

Sending the configuration command Kh sets the read timeout. One to five ASCII digits in the range of 10 to 65,000 determine the number of 10ms increments of time that the 3700 will, once triggered, wait while attempting to read from a port. For example, if you wanted to set port 1 to timeout ends a read attempt after 100ms; you would send the following command: <Kh , , 0, 10>. Port 2 is not changed. The default setting is Mode_global 0 and Time_global 150 (150 x 10 msec = 1.5 seconds).

4–10

Configuring the SCANTEAM 3700

Image 36
Contents SCANTEAMR3700 CCD Disclaimer Patents Page Table of Contents Chapter Configuring the Scanteam Technical Specifications Serial Commands SummaryCCD Operation and Bar Code Scanning Ascii Conversion TableList of Tables Preface Intended AudienceOverview of the Technical Manual Related Publications Scanteam 3700 Description IntroductionScanteam 3700 Hardware Description Scanteam 3700 Software Description Interface PortNonvolatile Memory System Hardware Description Basic System OperationGeneral Characteristics Mechanical LayoutSingle Modular Input /Output/Power Port Audible IndicatorOperating Theory Mounting InsertsScanner/Host Communication Communications PortAsynchronous Serial Ascii Interface Hardwire Pinouts Auxiliary TriggerPower Requirements SET-UP and Installation General Preparation for UseSet-up Procedure for Evaluation Set-up ChecklistPreparation Set-up Procedure To setup serial communications to the scannerInstallation in the Host Instrument Top View Examples of Scanteam 3700 Barcode Orientation Pitch or Roll Scanner Illumination Clearance Configuring the Scanteam PreparationCommand Conventions Command FormatData string Operational Commands Reset Command aDisable Decoding Command DC4 Enable Decoding Command DC2Serial Trigger Command Save Configuration Command ZSave Configuration 3C 5A 3E Serial Trigger 3C 31 3EConfiguration Commands Set Communications Port Command Ka Command Ka baud, parity, stop bits, data bits ParameterOption Command Kd enable, data Parameter Code ID Value DefinitionScanteam 3700 Code ID Values Set Postamble Command Ke Command Ke enable, data ParameterSet Communications Protocol Command Kf Command Kf mode ParameterSet Trigger Mode Command Kg Command Kg modeglobal, modeport1, modeport2, # ParameterSet Read Timeout Command Kh Set No Read Message Command Kk Command Kk enable, data ParameterSet Votes Command Km Set Code 39 Command Kp Set UPC A/E, EAN 8/13 Command Ks Command Ks enable UPC, enable EAN ParameterSet Code 128 Command Kt Set Beeper Mode Command KuSet Scan Rate Command Kv Configuration Status Query Commands Default Operating Parameters Default Configuration SettingsDefault Configuration Symbology Settings SERVICE/TECHNICAL Assistance MaintenanceTroubleshooting Guide Troubleshooting HintsDiagnostic Procedure Troubleshooting ChecklistProblem Possible Cause Solution Scanteam 3700 Troubleshooting ChecklistObtaining Factory Service Product Service DepartmentService Under Warranty Out Of Warranty ServiceHelp Desk Scanteam 3700 Removal/Replacement Followed during installation ChapterService/Technical Assistance Scanning and Decoding a Bar Code Symbol CCD Operation and BAR Code Scanning aObjectives LED Reflector CCD SensorPeak Detectors and Comparator Microprocessor Timing SequenceCloser Look At Bar Code Symbols Bar Code Basics and Scanning TipsBar Code Scanning Technical Specifications Optical SpecificationsScanner Performance MRD minimum reflective difference 37.5% Skew AnglePitch Angle Light Source IlluminationElectrical Specifications Environmental Specifications Mechanical SpecificationsBar Code Symbol Specifications Scanteam 3700 Interface Cable To Host SystemSerial Commands Summary Command FormatCommandFormat Individual Configuration QueryKa? Kd?Default Configuration Settings Configuration CommandDefault SettingsSymbology Configuration Default Settings Serial Commands Summary Ascii Conversion Table Control Function DefinitionsAscii Conversion Table Scanteam 3700 Glossary Glossary-2 Glossary-3 Glossary-4 Glossary-5 Glossary-6 Index Scanteam Index-3 Index-4 Scanteam 3700 Limited Warranty Jordan Road Box Skaneateles Falls, New York 3700/TM Rev E