Agilent Technologies E6701A manual View the log data on the PC, Save the Log to the PC

Page 19

4.For this tutorial, select GMM/SM and GSM_L3. Make sure the other observation points are NOT selected.

5.The Observation Points correspond to seven protocol layers you can log. Logging data will only be collected for layers that have been selected prior to the start of logging.

L1_Interface

Raw information between the DUT and Test Set

RLC/MAC

Access and control messages

LLC

Unnumbered and supervisory messages

GMM/SM

Attach, detach, security, and PDP context messages

GSM_L3

Common control channel messages

SNDCP

Network-level messages

IP

IP datagram data and control fields

6.Buffer Operation: for this tutorial, select Linear.

The Linear buffer stops the log when the logging memory location is full.

The Cyclic buffer writes over the oldest data once it is full.

7.Scroll to the bottom of the page.

8.Enter a User Comment if desired to annotate the log. This must be alphanumeric characters only, up to 74 characters in length.

9.Select OK to save the setup and return to the main logging page.

10.Select the Start Logging icon.

11.Turn on power to the phone (to perform a switch-on and attach).

Motorola Timeport Example:

The phone should indicate service signal-strength bars (in the upper-left corner of the display) within about 1 minute.

It should also indicate the numbers “001-01” in the center of the display (this is the default test network number stored in the Test Set).

12.On the Test Set, in the bottom center of the display, the Active Cell box should show Attached.

13.Select the Stop Logging icon. Note that no data is displayed until the log is stopped. The Test Set has captured the log data in its internal memory.

View the log data on the PC

1.Look in the PC’s upper window using the scroll bar to see the Message Summary. If you click on one of the message summaries, a hyperlink will launch in the lower window to show you the detail for that message.

2.Look in the lower window using the scroll bar to see the Message Details.

Notice the time stamp and message number at the front of each message. Refer to the E6701A Lab Application User’s Guide for more details on logged messages.

Save the Log to the PC

To save this log data in the PC, follow these steps:

1. At the top left corner of the Internet Explorer window, select File, then Save As.

 Copyright 2002 Agilent Technologies.

10001810-E6701A Tutorial LAN Connection

All Rights Reserved.

Page 19 of 32

Rev. 1.2; Feb. 14, 2002

Image 19
Contents Page Table of Contents Appendix B Troubleshooting Appendix C. Glossary Appendix a Installing a WAP Gateway into a PCIntroduction About This TutorialEquipment Required for this Tutorial Normal LAN Cable Quantity Setup Test Set Setup Network SetupLAN Addresses Set the RF In/Out Amplitude Offset Set the Date and Time OptionalSet the Cell and Call Parameters Windows 98 and Dhcp Address PC SetupWindows 98 and Static Address Page Windows 2000 and Dhcp Address Set up NetworkWindows 2000 and Static Address Determine PC IP address Verify PC Settings Motorola Timeport phone example Set Up Your Wireless DeviceOn the WAP Settings menu, scroll to 1Primary IP Verify that Your Device Will Perform an AttachSet WAPlite Gateway Address in Your Device On the WAP Settings menu, scroll to 2Primary PortSet the Device’s Primary Port Set Data Bearer to GprsSetup is now complete Product Feature Protocol Logging Product Feature Screen Image CaptureHere are some important points about the log data Save the Log to the PC View the log data on the PCView the saved log Log Data Using the Test Set’s Front PanelPinging the DUT the phone or wireless device Product Feature PingPinging an Alternate Device such as your PC on the LAN If your LAN does not use a firewall to access the Internet Product Feature Data ChannelMotorola Example Scroll to http//local/waplite.wmlMotorola Timeport Example Product Feature Additional Connection Parameters To load the WAPlite software on your PC Appendix a Installing a WAP Gateway into a PCInfinite WAPLite Gateway by Captaris Inc To download the WAPlite softwareIf the wireless device will not attach, check that Appendix B TroubleshootingIf Bler mode measurements have problems If Internet Explorer cannot access the 8960 Welcome Etsi Test Mode measurements have problemsBER measurement errors Phone can’t access browse real Internet sitesPhone won’t transmit IP datagrams in Data Channel functions Page Appendix C. Glossary