Canon RE-350 manual RTS flow control timing chart, Mark Duration period Min Max, 44ms 300ms

Page 59

6.2 RTS flow control timing chart

The PC can suspend the transmission of the “C-response Packet” data for the period less than 1000 milliseconds from the RE-350by setting the RTS signal to OFF.

PC SIDE SIGNAL

 

 

 

T6

RTS

 

1

3

CTS

 

 

 

TXD

Command

 

T7

 

T5

 

 

 

RXD

C- Response

2 4

Mark

Duration period

 

Min.

Max.

T5 (9600bps)

3.44ms

300ms

(2400bps)

4.58ms

300ms

T6

--

1000ms

T7

--

200ms

1If the PC wants to suspend the transmission of the “C-response Packet” data stream on its half, the PC should set the RTS line to OFF.

2Once the RE-350recognizes the RTS of the PC is OFF, the RE-350will hold the transmission of the “C-response Packet” data after the remaining maximum 3 byte data in its communication buffer is transmitted. The application software on the PC can recognize if the RE-350is operating properly. The time period required for transmitting the remaining data in the buffer will depend upon the communication speed of the RS232C I/F.

3When the PC wants to resume the transmission, the PC should set RTS to ON immediately. The longest time period to be able to hold the transmission is 1000 milliseconds. If the PC fails to set the RTS line to ON in 1000 milliseconds, the RE-350will stop the sequence to send the “C-response Packet”. The PC can not receive the remaining C-response data any more.

4Once the RE-350recognizes the RTS signal comes back to ON within 1000 milliseconds, the RE-350will resume the transmission of the “C-response Packet” data.

59

Image 59
Contents RE-350 Control Interface Command Index Operations and Functions Panel DiagramsButton Description of the Front and Rear Panel FunctionsKey Function Detail Button Power supply Control Knob Digital SwitchesMain power is turned on SCp180 Operations RE-350s Condition and LED Status Operational Procedures OverviewOperations RS-232 LED remains on RE-350will operate according to the givenAvailable Front panel PC and the RS-232 LED will blinkSetting the center target value for Auto-White Balance Mode MeaningAdjusting the value of each control knob LineRequest OFF Line Function of RS-232 buttonLineRequest on Line Status Default Setting of Each Function After Power-OnAuto-white balance was adjusted successfully Meaning of Each LED StatusParameter Control ProtocolPacket Discrimination Code Assignment ID numberEvent Data Packet Format PC RE-350 Data Packet ConfigurationControl Command Packet Format PC RE-350 Response Data Packet Format PC RE-350Response Flow Pattern 2 Event Packet Flow Pattern Operational Flow ChartFlow Pattern 1 Command Communication Flow Pattern CommandResponse data Flow Pattern 3 Notification Mode Flow PatternFP Data Packet Camera Control Commands Table Control Command and Packet TablesLED System Control Commands TableResponse Data Packet Table System error responseMode error response Time out error responseRequest on line Event Data Packet TableEvent Description Byte Header ID # ParaLDenied AcceptedParaH ParaL FP Data Packet TableFP data description Byte $10 $ 0 Description Command Data PacketZoom Commands Wide Zoom Constant SpeedTele Zoom Step Tele Zoom Constant SpeedZoom Stop Zoom Access To PositionZoom Request Position Zoom Tele Constant Speed w/ AF † Zoom Wide Constant Speed w/ AF †Focus Near Constant Speed Focus CommandsOP AF Operate Auto Focus $10 $ 1 $ 0Function Focus Near StepFocus Far Constant Speed Focus Stop Focus Far Step$10 $ 2 $ 0 White Balance CommandsWhite Balance Manual Access To OP AWB Operate Auto-White Balance$10 $ 2A $ 0 Exposure CommandsExposure Manual Access To Detail Manual Access To Detail Commands$10 $ 3 $ 0 Positive/Negative CommandsSet Positive Mode † Set Negative Mode †$10 $ 4 $ 0 Color/B&W CommandsSet Color Mode Set B/W Mode$10 $ 5 $ 0 System Control CommandsDetailed System Control Commands Select Document VideoDocument Lights Off † Document Lights On †Backlight Off † Backlight On †$10 $ 6 $ 0 LED NormalLED On $10 $ 6 a $ 0 LED Off$10 $ 6 B $ 0 LED Blink$10 $ 7 $ 0 OFF-LINE ModeON-LINE Mode Notification Mode $10 $ 8 $ 0 Maximum operation time msec 100 millisecond Command formatStatus of Group a Video C/M= on Status of Group BInput Function To get the status data of Group B from the RE-350Backlight Status of the ButtonsFocus Input RS-232 AutoParameter of the C-response Packet MSH $00 fixed Status of WB control knobStatus of EXP control knob Function Gets the position data of the Detail control knob Status of Detail control knobStatus of AWB result Parameter of the C-response Packet Name of the EquipmentVersion of ROM $10 $F F $ 0 Set to Special Mode$10 $F F $ F F Set to Service ModeRS-232C Control Cable Pin Assignment Control Cable Pin AssignmentsRE-350 RE-350 Control Cable Pin Assignments Pin Dsub RS-232CPage Mark Duration period Min Max 300ms 200ms 10ms T4-T2 Timing ChartsRTS/CTS Flow Control Timing Chart Command58ms 300ms 1000ms 200ms RTS flow control timing chartMark Duration period Min Max 44ms 300msAppendix a Disk Information RE-350 Software Developers Disk InstallationRE-350 SDK File Definitions RE-350 Demo Application \APPClose communications with the RE-350 Execute an RE-350commandRE-350 16bit Sample Code \16BITSDK Initialize communications with the RE-350RE-350 32bit Sample Code \32BITSDK Execute an RE-350 command