JAI CV-M4+, M7+ CL, CV-M7+, M4+ CL operation manual Frame-delay read out Mode with multiple exposure

Page 21

CV-M4+/M4+CL, CV-M7+/M7+CL

6.8. Frame-delay read out Mode with multiple exposure.

Multiple exposures is possible in frame delay read out mode. The pre selected exposure starts from the leading edge of the trigger pulse. It can be the 10 steps in normal or 1056 steps in programmable. SM=0 or SM=1. The resulting video is read out at the trailing edge of the trigger. This mode will operate with full and partial scanning and with binning (M4+ and M4+CL only). Partial scanning and binning in combinations is not allowed.

If smearless read out (SL=1) is used, the first exposure starts with a delay 133H (5.23 msec.) after the trigger.

Input of the multiple shutter trigger can be applied >1H after the exposure time (EEN low). An EEN pulse will indicate the active accumulation time, and a FEN pulse indicates that the resulting video is read out.

To use this mode:

 

Set function:

Trigger mode “Frame delay read out”

TR = 3

 

Multiple exposure

ML = 1

 

“H synchronous accumulation”

HC=0, HC=1

 

Polarity and other functions

 

Input:

Ext. trigger to LVDS/Camera Link or pin 10 on 12-pin connector.

 

Ext. Multi shutter to LVDS/Camera Link or pin 11 on 12-pin connector.

Important notes on using this mode.

The duration of the trigger should be low >2H after the end of the last exposure.

The number of exposures in multiple shutter should be 6.

A new trigger must not be applied before FEN is high.

Fig. 24. Multiple exposure, normal and smearless

- 20 -

Image 21
Contents CV-M4+/M4+CL CV-M7+/M7+CL CV-M4+/M4+CL, CV-M7+/M7+CL Main Features GeneralStandard Composition Locations Locations and FunctionsPin Multi-connector DC-IN/RS232C Digital Output Connector for EIA-644 LvdsPin Assignment Digital Output Connector for Camera Link Video signal, LEN, FENDval and EEN Video output Input and Output CircuitsTrigger input Multi Shutter input Composite Sync outputMUX Camera Link interfaceCV-M4+Block Diagram Functions and OperationsBasic functions Binning Color coding from a Primary RGB CCD Bayer filter Color versions CV-M7+and CV-M7+CLRestart Continuous Trigger mode. RCT Output of Timing SignalsPolarity and other functions Continuous Operation Non triggeredVertical timing details, full frame Vertical timing details with 1/2 partial scanning Accumulation latency time for H a-synchronous trigger External Trigger ModesImportant notes on using this mode Edge Pre-select ModeEdge Pre-select, smearless Pulse Width Pulse Width Control ModeFrame Delay Frame-delay read out ModeMultiple exposure, normal and smearless Frame-delay read out Mode with multiple exposureGain and set-up Other FunctionsMode setting SW1 on rear Configuring the CameraMode setting SW301 inside Internal Switch and Jumper SettingsRS-232C control Timing and shutter related commands CV-M4+command listCamera Control Tool for CV-M4+ External Appearance and Dimensions SpecificationsSpectral sensitivity Specifications Specification tablePrecautions Typical CCD CharacteristicsAppendix Camera Link test points Camera Link Test pointsUsers Record Users Mode Settings Users Modifications