Teledyne 3020T operating instructions Examples, Analog Output Connections

Page 27

Trace Oxygen Analyzer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Installation 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3-5: Analog Output Connections

The outputs are:

0–1 V dc % of Range: Voltage rises linearly with increasing oxygen, from 0 V at 0 ppm to 1 V at full scale ppm. (Full scale = 100% of programmable range.)

0–1 V dc Range ID: 0.25 V = Low Range, 0.5 V = Medium Range, 0.75 V = High Range, 1 V = Air Cal Range.

4–20 mA dc % Range: Current increases linearly with increasing oxygen, from 4 mA at 0 ppm to 20 mA at full scale ppm. (Full scale = 100% of programmable range.)

4–20 mA dc Range ID: 8 mA = Low Range, 12 mA = Medium Range, 16 mA = High Range, 20 mA = Air Cal Range.

Examples:

The analog output signal has a voltage which depends on the oxygen concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode.

The signal output for concentration is linear over the currently selected analysis range. For example, if the analyzer is set on a range that was defined as 0–100 ppm O 2, then the output would be as shown in Table 3-1.

Teledyne Analytical Instruments

3-5

Image 27
Contents Teledyne Analytical Instruments Model 3020TCopyright 1999 Teledyne Analytical Instruments Model 3020T3020T-I Trace Oxygen AnalyzerIii Specific Model InformationTable of Contents Introduction InstallationOperational Theory Operation MaintenanceAppendix Typical Applications Main Features of the AnalyzerTrace Oxygen Analyzer Introduction OverviewModel 3020T Model DesignationsTrace Oxygen Analyzer Introduction Operator Interface Model 3020T Controls, Indicators, and Connectors1 UP/DOWN Switch ESCAPE/ENTER SwitchPWD Displays Electrical Connector PanelRecognizing Difference Between LCD Equipment InterfaceRemote Valves RS-232 PortGas Connector Panel Remote SensorNetwork I/O Optional Operational Theory Introduction Micro-Fuel Cell SensorPrinciples of Operation Anatomy of a Micro-Fuel Cell Operational Theory+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSample System Characteristic Input/Output Curve for a Micro-Fuel CellSpan Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3020T Electronics Temperature Control Trace Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Front View of the Model 3020T Simplified Required Front Door Clearance Electrical ConnectionsPrimary Input Power Fuse InstallationAnalog Outputs Analog Output Connections ExamplesRange Voltage Current mA Alarm RelaysTypes of Relay Contacts Digital Remote Cal InputsZero RS-232 Sig RS-232 Pin Purpose Range ID RelaysNetwork I/O 8 RS-232 PortCommand Description Parameter SettingRemote Sensor and Solenoid Valves Remote Solenoid Return Connector Pinouts Installing the Micro-Fuel CellGas Connections Gas Connector Panel Testing the System Using the Controls Trace Oxygen Analyzer OperationAuto Ranging on Analysis Mode Setup ModeMode/Function Selection Trace Oxygen Analyzer Data Entry EnterEscape Trace Oxygen Analyzer Operation AUTO-CALFunction PWD Password FunctionEntering the Password Only one password can be definedInstalling or Changing the Password Version Screen Logout FunctionZero and Span Functions Trace Oxygen Analyzer Operation SELF-TESTFunctionZero Cal Auto Mode ZeroingManual Mode Zeroing Auto Mode Spanning Cell FailureSpan Cal Manual Mode Spanning Installation Alarms FunctionAL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Standby Function Contrast FunctionAnalysis Mode Operation Trace Oxygen Analyzer Maintenance Routine MaintenanceMajor Internal Components Major Internal Components Cell ReplacementWhen to Replace a Cell Storing and Handling Replacement CellsRemoving the Micro-Fuel Cell Exploded View of Cell Block and Micro-Fuel Cell Installing a New Micro-Fuel Cell Fuse Replacement Cell WarrantyL-2C cell is not designed for applications where CO2 is a Removing Fuse Cap and Fuse from Holder System Self Diagnostic TestPower AnalogPreamp Maintenance Trace Oxygen Analyzer Appendix SpecificationsOperating Temperature 0-50 C Qty Part Number Description Recommended 2-Year Spare Parts ListDrawing List Series Analyzers Restrictor KIT Conversions