Teledyne 3020T operating instructions Gas Connections

Page 34

3 Installation

Model 3020T

 

 

 

 

 

 

 

 

 

When the micro-Fuel Cell needs to be installed or replaced, follow the procedures in chapter 5, Maintenance, for removing and installing cells.

3.5Gas Connections

Before using this instrument, it should be determined if the unit will be used for pressurized service or vacuum service and low pressure applica- tions. Inspect the restrictor kit that came with the unit. The kit consist of two restrictors and a union for 1/4” diameter tubing. Notice that the two 1 3/4” long, 1/4” diameter tubing are restrictors. It has an open end and a closed end with a small circular orifice. The restrictor without the blue sticker is for ;ow pressure and vacuum service. For high pressure (2 to 50 psig) applica- tions, use the restrictor that has a blue sticker on the body.

For pressurized service, use the restrictor without the blue dot and union from the restrictor kit and attach it to the Sample In port. The small circular orifice should face away from the back of the unit (against the direction of gas flow). Use the restrictor without the blue dot sticker in the same manner for low pressure applications (less than 5 psig).

For vacuum service (5-10 in Hg), use the restrictor without the blue dot sticker and union but attach it to the Exhaust Out port. The small circular orifice should face toward the back of the unit (against the direction of gas flow).

Remove the blue sticker from the restrictor before using.

WARNING: Operating the unit without restrictors can cause damage to t the micro-fuel cell.

Figure 3-9 is an illustration of the Gas Connector Panel. Optional gas connections are shown in a shaded block.

The unit is manufactured with 1/4 inch tube fittings only. Adapters must be used for metric tubing. (At least 6 mm is recommended.)

For a safe connection:

1.Insert the tube into the tube fitting, and finger-tighten the nut until the tubing cannot be rotated freely, by hand, in the fitting. (This may require an additional 1/8 turn beyond finger-tight.)

2.Hold the fitting body steady with a backup wrench, and with another wrench rotate the nut another 11/4 turns.

3-12

Teledyne Analytical Instruments

Image 34
Contents Model 3020T Teledyne Analytical InstrumentsModel 3020T Copyright 1999 Teledyne Analytical InstrumentsSpecific Model Information Trace Oxygen AnalyzerIii 3020T-IInstallation Table of Contents IntroductionOperational Theory Maintenance OperationAppendix Overview Main Features of the AnalyzerTrace Oxygen Analyzer Introduction Typical ApplicationsModel Designations Model 3020TModel 3020T Controls, Indicators, and Connectors Trace Oxygen Analyzer Introduction Operator InterfaceESCAPE/ENTER Switch 1 UP/DOWN SwitchPWD Equipment Interface Electrical Connector PanelRecognizing Difference Between LCD DisplaysRS-232 Port Remote ValvesRemote Sensor Gas Connector PanelNetwork I/O Optional Micro-Fuel Cell Sensor Operational Theory IntroductionPrinciples of Operation Operational Theory Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H 2O + 4e → 4OHCalibration Characteristics Effect of PressureCharacteristic Input/Output Curve for a Micro-Fuel Cell Sample SystemElectronics and Signal Processing Span Zero SampleBlock Diagram of the Model 3020T Electronics Temperature Control Unpacking the Analyzer Trace Oxygen Analyzer InstallationMounting the Analyzer Front View of the Model 3020T Simplified Electrical Connections Required Front Door ClearanceFuse Installation Primary Input PowerAnalog Outputs Examples Analog Output ConnectionsAlarm Relays Range Voltage Current mADigital Remote Cal Inputs Types of Relay ContactsZero 8 RS-232 Port Range ID RelaysNetwork I/O RS-232 Sig RS-232 Pin PurposeParameter Setting Command DescriptionRemote Sensor and Solenoid Valves Installing the Micro-Fuel Cell Remote Solenoid Return Connector PinoutsGas Connections Gas Connector Panel Testing the System Trace Oxygen Analyzer Operation Using the ControlsAuto Ranging on Setup Mode Analysis ModeMode/Function Selection Trace Oxygen Analyzer Enter Data EntryEscape PWD Password Function Trace Oxygen Analyzer Operation AUTO-CALFunctionOnly one password can be defined Entering the PasswordInstalling or Changing the Password Logout Function Version ScreenTrace Oxygen Analyzer Operation SELF-TESTFunction Zero and Span FunctionsAuto Mode Zeroing Zero CalManual Mode Zeroing Cell Failure Auto Mode SpanningSpan Cal Manual Mode Spanning Alarms Function InstallationAL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionFixed Range Analysis Contrast Function Standby FunctionAnalysis Mode Operation Routine Maintenance Trace Oxygen Analyzer MaintenanceMajor Internal Components Cell Replacement Major Internal ComponentsStoring and Handling Replacement Cells When to Replace a CellRemoving the Micro-Fuel Cell Exploded View of Cell Block and Micro-Fuel Cell Installing a New Micro-Fuel Cell Cell Warranty Fuse ReplacementL-2C cell is not designed for applications where CO2 is a System Self Diagnostic Test Removing Fuse Cap and Fuse from HolderAnalog PowerPreamp Maintenance Specifications Trace Oxygen Analyzer AppendixOperating Temperature 0-50 C Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List Series Analyzers Restrictor KIT Conversions