Teledyne 3020T operating instructions Conversions

Page 73

Trace Oxygen Analyzer

Appendix

 

 

 

 

 

 

 

 

 

BY-PASS:

To improve the system response, a by-pass can be added to increase the sample flow rate to the analyzer by a factor of ten. A by-pass provides a sample flow path around the analyzer of 2 - 18 SCFH. typically.

CALIBRATION GAS:

3000 series analyzer requirements for units with Auto-Cal options. The customer must supply a control valves (or restrictors) for any SPAN or ZERO gas source which is attached to the Auto-Cal ports. The valve should be adjusted to the same flow rate as the sample gas .

When restrictors are used, the gas pressure must be adjusted to achieve the proper flow rate.

OPERATION WITHOUT A RESTRICTOR DEVICE:

Operation without a restrictor device is not recommend as mentioned above. A

3020T without any flow restrictor device was tested on 11-19-97. This results in a flow rate of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the standard sample sys.

REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on flow rates at

various pressures.

 

 

 

TAI PART NUMBERS

 

 

RESTRICTOR KIT:

A68729

 

UNION (SS)

 

U11

 

LP. RESTRICTOR

R2323 ( LOW PRESSURE / VAC. SERVICE )

STD.. RESTRICTOR

R2324

BLUE DOT

NUT

 

N73

 

FERRULE

 

F73

 

FERRULE

 

F74

BOTH FERRULES ARE

 

 

 

REQUIRED

CONVERSIONS:

 

 

1 PSI

=

2.04 INCHES OF MERCURY (in. Hg.)

1 SCFH =

0.476 SLPM

 

Teledyne Analytical Instruments

A-7

Image 73
Contents Teledyne Analytical Instruments Model 3020TCopyright 1999 Teledyne Analytical Instruments Model 3020TIii Trace Oxygen AnalyzerSpecific Model Information 3020T-IInstallation Table of Contents IntroductionOperational Theory Operation MaintenanceAppendix Trace Oxygen Analyzer Introduction Main Features of the AnalyzerOverview Typical ApplicationsModel 3020T Model DesignationsTrace Oxygen Analyzer Introduction Operator Interface Model 3020T Controls, Indicators, and ConnectorsESCAPE/ENTER Switch 1 UP/DOWN SwitchPWD Recognizing Difference Between LCD Electrical Connector PanelEquipment Interface DisplaysRemote Valves RS-232 PortRemote Sensor Gas Connector PanelNetwork I/O Optional Micro-Fuel Cell Sensor Operational Theory IntroductionPrinciples of Operation Anatomy of a Micro-Fuel Cell Operational Theory+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSample System Characteristic Input/Output Curve for a Micro-Fuel CellSpan Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3020T Electronics Temperature Control Unpacking the Analyzer Trace Oxygen Analyzer InstallationMounting the Analyzer Front View of the Model 3020T Simplified Required Front Door Clearance Electrical ConnectionsFuse Installation Primary Input PowerAnalog Outputs Analog Output Connections ExamplesRange Voltage Current mA Alarm RelaysTypes of Relay Contacts Digital Remote Cal InputsZero Network I/O Range ID Relays8 RS-232 Port RS-232 Sig RS-232 Pin PurposeParameter Setting Command DescriptionRemote Sensor and Solenoid Valves Remote Solenoid Return Connector Pinouts Installing the Micro-Fuel CellGas Connections Gas Connector Panel Testing the System Trace Oxygen Analyzer Operation Using the ControlsAuto Ranging on Setup Mode Analysis ModeMode/Function Selection Trace Oxygen Analyzer Enter Data EntryEscape Trace Oxygen Analyzer Operation AUTO-CALFunction PWD Password FunctionEntering the Password Only one password can be definedInstalling or Changing the Password Version Screen Logout FunctionZero and Span Functions Trace Oxygen Analyzer Operation SELF-TESTFunctionZero Cal Auto Mode ZeroingManual Mode Zeroing Cell Failure Auto Mode SpanningSpan Cal Manual Mode Spanning Installation Alarms FunctionAL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Standby Function Contrast FunctionAnalysis Mode Operation Routine Maintenance Trace Oxygen Analyzer MaintenanceMajor Internal Components Major Internal Components Cell ReplacementWhen to Replace a Cell Storing and Handling Replacement CellsRemoving the Micro-Fuel Cell Exploded View of Cell Block and Micro-Fuel Cell Installing a New Micro-Fuel Cell Cell Warranty Fuse ReplacementL-2C cell is not designed for applications where CO2 is a Removing Fuse Cap and Fuse from Holder System Self Diagnostic TestAnalog PowerPreamp Maintenance Trace Oxygen Analyzer Appendix SpecificationsOperating Temperature 0-50 C Qty Part Number Description Recommended 2-Year Spare Parts ListDrawing List Series Analyzers Restrictor KIT Conversions