1 Introduction | Model 3020T |
| | |
| | |
| | |
•Advanced Micro-Fuel Cell, redesigned for trace analysis, has an expected life of one year.
•Versatile analysis over a wide range of applications.
•Microprocessor based electronics: 8-bit CMOS microprocessor with 32 kB RAM and 128 kB ROM.
•Three user definable output ranges (from 0-10 ppm through 0- 250,000 ppm) allow best match to users process and equipment.
•Air-calibration range for convenient spanning at 20.9 %.
•Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest.
•Two adjustable concentration alarms and a system failure alarm.
•Extensive self-diagnostic testing, at startup and on demand, with continuous power-supply monitoring.
•RS-232 serial digital port for use with a computer or other digital communication device.
•Four analog outputs: two for measurement (0–1 V dc and Isolated 4–20 mA dc) and two for range identification.
1.4Model Designations
3020T: Standard model.
3020T-C:In addition to all standard features, this model also has separate ports for zero and span gases, and built-in control valves. The internal valves are entirely under the control of the 3020T electronics, to automatically switch between gases in synchronization with the analyzer’s operations
3020T-F:Flame arrestors for Groups C and D installed.
3020T-G:Flame arrestors for Groups C and D, & -C option, installed.
3020T-H:Flame arrestors for Group B (hydrogen) installed.
3020T-I:Flame arrestors for Group B, & -C option, installed.
All of the above options are available in combination.