Teledyne 3020T operating instructions Gas Connector Panel

Page 35

Trace Oxygen Analyzer

 

 

 

 

 

 

 

 

 

 

 

 

Installation 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3-9: Gas Connector Panel

SAMPLE IN: In the standard model, gas connections are made at the SAMPLE IN and EXHAUST OUT connections. Calibration gases must be Tee'd into the Sample inlet with appropriate valves.

The gas pressure in should be reasonably regulated. Pressures between 3 and 40 psig are acceptable as long as the pressure, once established, will keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4 SLPM). Exact figures will depend on your process.

If greater flow is required for improved response time, install a bypass in the sampling system upstream of the analyzer input.

Note: If the unit is for vacuum service, the above numbers apply instead to the vacuum at the EXHAUST OUT connector, de- scribed below, with minus signs before the pressure readings.

ZERO IN and SPAN IN: These are additional ports for inputting span gas and zero gas. There are electrically operated valves inside for automatic switching between sample and calibration gases. These valves are com- pletely under control of the 3020T Electronics. They can be externally controlled only indirectly through the Remote Cal Inputs, described below.

Pressure, flow, and safety considerations are the same as prescribed for the SAMPLE IN inlet, above.

EXHAUST OUT: Exhaust connections must be consistent with the hazard level of the constituent gases. Check Local, State, and Federal laws, and ensure that the exhaust stream vents to an appropriately controlled area if required.

Teledyne Analytical Instruments

3-13

Image 35
Contents Teledyne Analytical Instruments Model 3020TCopyright 1999 Teledyne Analytical Instruments Model 3020T3020T-I Trace Oxygen AnalyzerIii Specific Model InformationOperational Theory Table of Contents IntroductionInstallation Operation MaintenanceAppendix Typical Applications Main Features of the AnalyzerTrace Oxygen Analyzer Introduction OverviewModel 3020T Model DesignationsTrace Oxygen Analyzer Introduction Operator Interface Model 3020T Controls, Indicators, and ConnectorsPWD 1 UP/DOWN SwitchESCAPE/ENTER Switch Displays Electrical Connector PanelRecognizing Difference Between LCD Equipment InterfaceRemote Valves RS-232 PortNetwork I/O Gas Connector PanelRemote Sensor Optional Principles of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Anatomy of a Micro-Fuel Cell Operational Theory+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSample System Characteristic Input/Output Curve for a Micro-Fuel CellSpan Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3020T Electronics Temperature Control Mounting the Analyzer Trace Oxygen Analyzer InstallationUnpacking the Analyzer Front View of the Model 3020T Simplified Required Front Door Clearance Electrical ConnectionsAnalog Outputs Primary Input PowerFuse Installation Analog Output Connections ExamplesRange Voltage Current mA Alarm RelaysTypes of Relay Contacts Digital Remote Cal InputsZero RS-232 Sig RS-232 Pin Purpose Range ID RelaysNetwork I/O 8 RS-232 PortRemote Sensor and Solenoid Valves Command DescriptionParameter Setting Remote Solenoid Return Connector Pinouts Installing the Micro-Fuel CellGas Connections Gas Connector Panel Testing the System Auto Ranging on Using the ControlsTrace Oxygen Analyzer Operation Mode/Function Selection Analysis ModeSetup Mode Trace Oxygen Analyzer Escape Data EntryEnter Trace Oxygen Analyzer Operation AUTO-CALFunction PWD Password FunctionEntering the Password Only one password can be definedInstalling or Changing the Password Version Screen Logout FunctionZero and Span Functions Trace Oxygen Analyzer Operation SELF-TESTFunctionZero Cal Auto Mode ZeroingManual Mode Zeroing Span Cal Auto Mode SpanningCell Failure Manual Mode Spanning Installation Alarms FunctionAL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Standby Function Contrast FunctionAnalysis Mode Operation Major Internal Components Trace Oxygen Analyzer MaintenanceRoutine Maintenance Major Internal Components Cell ReplacementWhen to Replace a Cell Storing and Handling Replacement CellsRemoving the Micro-Fuel Cell Exploded View of Cell Block and Micro-Fuel Cell Installing a New Micro-Fuel Cell L-2C cell is not designed for applications where CO2 is a Fuse ReplacementCell Warranty Removing Fuse Cap and Fuse from Holder System Self Diagnostic TestPreamp PowerAnalog Maintenance Trace Oxygen Analyzer Appendix SpecificationsOperating Temperature 0-50 C Qty Part Number Description Recommended 2-Year Spare Parts ListDrawing List Series Analyzers Restrictor KIT Conversions