Trace Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
4.3The AUTO-CAL Function
When proper automatic valving is connected (see chapter 3, installa- tion), the Analyzer can cycle itself through a sequence of steps that automati- cally zero and span the instrument.
Note: If you require highly accurate
Model 3020T is accurate to
To setup an
Scroll to
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
If SPAN (or ZERO) is not between the arrows, scroll with DOWN/UP
control to SPAN (or ZERO), then ENTER. (You won’t be able to set OFF to ON if a zero interval is entered.) A Span Every ... (or Zero Every ...)
screen appears.
Span Every 0 d
Start 0 h from now
Use DOWN/UP control to set an interval value, and ENTER. Then use DOWN/UP to set a
To turn ON the Span and/or Zero cycles (to activate
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
When the Span/Zero values screen appears, use the scroll DOWN to blink the OFF/ON field of the SPAN (or ZERO) function. Use DOWN/UP to set the OFF/ON field to ON. You can now turn these fields ON because there is a nonzero span interval defined.
4.4The PWD (Password) Function
Security can be established by choosing a 5 digit password from the standard ASCII character set. If you decide NOT to employ password security, use the default password TETAI. This password will be displayed automatically by the microprocessor. The operator just ENTERs it to be allowed total access to the instrument’s features.
Teledyne Analytical Instruments |