Teledyne 3020T operating instructions Fuse Replacement, Cell Warranty

Page 63

Trace Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

5.3.5 Cell Warranty

The Class L-2C Micro-Fuel cell is used in the Model 3020T. This cell is a long life cell and is warranted for 1 year from the date of shipment. If any other cell is supplied with your instrument, check the front of this manual for any special information applying to your cell.

With regard to spare cells, warranty period begins on the date of ship- ment. The customer should purchase only one spare cell (per section 5.3.1). Do not attempt to stockpile spare cells.

The L-2C cell is not designed for applications where CO2 is a

major component in the sample, however concentrations of 1,000 ppm or less will not adversely effect the cell performance. Consult Analytic Instru- ments for available options for either intermittent or continuous CO2 expo- sure.

If a cell was working satisfactorily, but ceases to function before the warranty period expires, the customer will receive credit toward the purchase of a new cell.

If you have a warranty claim, you must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty work- manship or material, the cell will be replaced at no cost to you.

Note: Evidence of damage due to tampering or mishandling will render the cell warranty null and void.

5.4Fuse Replacement

The 3020T requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses. The fuses are located inside the explosion proof housing on the Electrical Connector Panel, as shown in Figure 5-2. To replace a fuse:

1. Disconnect the Unit from its power source.

2.Place a small screwdriver in the notch in the fuse holder cap, push in, and rotate 1/4 turn. The cap will pop out a few millimeters. Pull out the fuse holder cap and fuse, as shown in Figure 5-3.

Teledyne Analytical Instruments

5-7

Image 63
Contents Teledyne Analytical Instruments Model 3020TCopyright 1999 Teledyne Analytical Instruments Model 3020T3020T-I Trace Oxygen AnalyzerIii Specific Model InformationTable of Contents Introduction InstallationOperational Theory Operation MaintenanceAppendix Typical Applications Main Features of the AnalyzerTrace Oxygen Analyzer Introduction OverviewModel 3020T Model DesignationsTrace Oxygen Analyzer Introduction Operator Interface Model 3020T Controls, Indicators, and Connectors1 UP/DOWN Switch ESCAPE/ENTER SwitchPWD Displays Electrical Connector PanelRecognizing Difference Between LCD Equipment InterfaceRemote Valves RS-232 PortGas Connector Panel Remote SensorNetwork I/O Optional Operational Theory Introduction Micro-Fuel Cell SensorPrinciples of Operation Anatomy of a Micro-Fuel Cell Operational Theory+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSample System Characteristic Input/Output Curve for a Micro-Fuel CellSpan Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3020T Electronics Temperature Control Trace Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Front View of the Model 3020T Simplified Required Front Door Clearance Electrical ConnectionsPrimary Input Power Fuse InstallationAnalog Outputs Analog Output Connections ExamplesRange Voltage Current mA Alarm RelaysTypes of Relay Contacts Digital Remote Cal InputsZero RS-232 Sig RS-232 Pin Purpose Range ID RelaysNetwork I/O 8 RS-232 PortCommand Description Parameter SettingRemote Sensor and Solenoid Valves Remote Solenoid Return Connector Pinouts Installing the Micro-Fuel CellGas Connections Gas Connector Panel Testing the System Using the Controls Trace Oxygen Analyzer OperationAuto Ranging on Analysis Mode Setup ModeMode/Function Selection Trace Oxygen Analyzer Data Entry EnterEscape Trace Oxygen Analyzer Operation AUTO-CALFunction PWD Password FunctionEntering the Password Only one password can be definedInstalling or Changing the Password Version Screen Logout FunctionZero and Span Functions Trace Oxygen Analyzer Operation SELF-TESTFunctionZero Cal Auto Mode ZeroingManual Mode Zeroing Auto Mode Spanning Cell FailureSpan Cal Manual Mode Spanning Installation Alarms FunctionAL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Standby Function Contrast FunctionAnalysis Mode Operation Trace Oxygen Analyzer Maintenance Routine MaintenanceMajor Internal Components Major Internal Components Cell ReplacementWhen to Replace a Cell Storing and Handling Replacement CellsRemoving the Micro-Fuel Cell Exploded View of Cell Block and Micro-Fuel Cell Installing a New Micro-Fuel Cell Fuse Replacement Cell WarrantyL-2C cell is not designed for applications where CO2 is a Removing Fuse Cap and Fuse from Holder System Self Diagnostic TestPower AnalogPreamp Maintenance Trace Oxygen Analyzer Appendix SpecificationsOperating Temperature 0-50 C Qty Part Number Description Recommended 2-Year Spare Parts ListDrawing List Series Analyzers Restrictor KIT Conversions