TECHNICAL SPECIFICATIONS
PRESSURE / FLOW CURVE
Note: As a result of manufacturing variations, the vent flow may vary from the values listed in the following table.
ENGLISH |
Pressure Flow (cm H2O) (L/min)
4 | 22 |
8 | 32 |
12 | 41 |
1648
2054
| 80 |
|
|
|
|
|
|
|
|
|
|
| 70 | 8 0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| 7 0 |
|
|
|
|
|
|
|
|
|
| 60 | 6 0 |
|
|
|
|
|
|
|
|
|
(L/min) | 50 | 5 0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| 4 0 |
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| |
Rate | 40 | 3 0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
Flow | 30 | 2 0 |
|
|
|
|
|
|
|
|
|
1 0 |
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| |
| 20 | 0 |
|
|
|
|
|
|
|
|
|
|
| 4 | 6 | 8 | 10 | 12 | 14 | 16 | 18 |
| 20 |
| 10 |
|
| M a s k P r e s s ur e ( c m H 2 0 ) |
|
|
| ||||
|
|
|
|
|
|
| |||||
| 0 |
|
|
|
|
|
|
|
|
|
|
|
| 4 | 6 | 8 | 10 |
| 12 | 14 | 16 | 18 | 20 |
Mask Pressure (cm H20)
FLOW GENERATOR SETTING
Select "Mir Full" as the mask option when using the ResMed Hospital Full Face Mask with ResMed flow generators that have mask setting options.
DEAD SPACE INFORMATION
Explanation: Dead space is the empty volume of the mask up to the elbow.
The dead space of the mask varies according to cushion sizes but is less than 317 mL.
ENVIRONMENTAL CONDITIONS
Operating Temperature: 5°C (41°F) to 40°C (104°F)
Operating Humidity:
SYMBOLS WHICH APPEAR ON THE DEVICE
Attention, consult accompanying documents.
ResMed Mask
S: Small; M: Medium; L: Large
REPLACEMENT PARTS
Part Item
No.
60700 ResMed Hospital Full Face Mask System, Small
60701 ResMed Hospital Full Face Mask System, Medium
60702 ResMed Hospital Full Face Mask System, Large
60710 Headgear (10pk.)