A&D TM-2430-13 instruction manual Example Programming the Monitor for Four 4 Periods, Off

Page 40

(C) Example: Programming the Monitor for Four (4) Periods

Suppose that the start time of the monitoring session is 09:00 and you would like the monitor to measure the patient's blood pressure every 30 minutes until 22:00, and from 22:00 to 23:00 you want the monitor OFF so that the patient may sleep, and from 23:00 to 06:00 you would like the patient's blood pressure measured once an hour, and finally from 06:00 you want the blood pressure measured every 15 minutes. Remember that the Time for Period One (1) also defines the End of the monitoring session. The end time for Period Four (4) will default to the Time for Period One (1), in this case 09:00.

PERIOD

TIME

INTERVAL

DURATION

1

09:00-22:00

30 minutes

13 hours

2

22:00-23:00

OFF

1 hours

3

23:00-06:00

1 hour

7 hours

4

06:00-09:00

15 minutes

3 hours

 

 

Total:

24 hours

The following is the recorder setup according to these instructions.

Note: Set Periods One (1), Two (2), Three (3) and Four (4). Leave the other Periods as “-”.

PERIOD 1

2

3

4

5

6

START INTERVAL

9

 

30 minutes

 

 

 

22

 

OFF

 

 

 

23

 

1 hour

 

 

 

6

 

15 minutes

 

 

 

-

-

 

 

 

-

-

Figure 25: Monitor programming example 3

39

Image 40
Contents TM-2430-13 Contents Troubleshooting Introduction About the Data Analysis SoftwareSymbols Monitoring Device ReviewInstalling the Software SafetySoftware Package Components System RequirementsConnecting the TM-2420/TM-2421 to the computer Getting StartedCommunication Between the Monitor and the Software Communication Between the Monitor and the SoftwareConnecting the TM-2430 series to the computer Connecting the UA-767PC to the computer Main Screen Description Analysis Intervals BP Mode ToolbarMethod Setting Description File Menu Menu DescriptionSave As DeleteFile Merge ConvertSelect the Merge the BP data files sub-menu Print Click OpenExit Saving a FileView Menu Summary Data display Summary Data Tab Definition Errors forBP Blood See 3-1 Measurement Errors StatusExcluding/Including BP Measurements Entering CommentsLinking to Trends Screen Systolic/Diastolic Blood Pressure BP TrendPulse Pulse Trend Mean Arterial Blood Pressure MAP TrendDelimitation, Scroll Bar and Linking Partial Analysis SleepBlood Pressure Trend BP Limits Scroll bar/LinkingScaling the Screen Viewing the TrendsTrend Display Positioning Trends Trend Page Setup dialog box Creating a New/Edit Trends TemplateBlood Pressure Trend Setup dialog box Correlation Plots display Systolic Blood Pressure Diastolic Blood PressureMAP, Mean Blood Arterial Pressure PulseCircadian Rhythm display HIB Analysis HIB analysis displayField Description Intervals and ConditionsData retrieval. See 2-7-1 Retrieving Data from the Monitor Physician ListRecorder Menu Select the Patient Information sub-menuSelect the Recorder menu Select the Retrieve data sub-menu Retrieving Data from the MonitorClick the Retrieve Data button Clearing The Monitor’s Memory Select the Recorder menuMonitoring session is performed at the same interval for TM-2420/TM-2421 Recorder Setup dialog box 3-2 TM-2430 Series Setup Description of each mode is as followsPeriod Start Interval Example Programming the Monitor for Three 3 Periods Monitor programming exampleExample Programming the Monitor for Four 4 Periods OFF3-4 UA-767PC Setup Blood Pressure Measurement Available only with the UA-767PCReport Menu Report dialog boxReport setup dialog box Creating a New Report TemplateSaving a Default Report Template for Printing Editing an Existing Report TemplateDeleting an Existing Report Template Printing an Existing Report TemplateOptions Menu Adding a New Physician to the List Editing an Existing Physician in the ListDeleting an Existing Physician in the List Trend Display See 2-6-3-3 Trend Display Positioning TrendsDefault Blood Pressure Settings dialog box Serial Port On-line HelpRecorder Type Measurement Errors 1 TM-2420/TM-2421 Error CodesBe erased Well as programmed intervalsBe taped to the arm with the supplied 2 TM-2430 Series Error Codes Communication Errors Communications Error dialog boxMissing Files Reports Are Not Printing With the Desired BP ModeSummary Data / Sleep Tab Data can not be Seen Printout Examples Summary data Full and partial Text data Summary data Awake and sleep Text data Blood pressure data Text data Trend graph Blood pressure and double product Graphical data Page Page Mini report Mini report Terms Related to Blood Pressure Terms Related to StatisticsTerms Related to Biological Rhythm Mesor95% confidence interval is expressed Terms Related to the Risk Management of High Blood Pressure Memo Memo Memo