A&D TM-2430-13 instruction manual Connecting the TM-2430 series to the computer

Page 8

Step 1 Connect the monitor to the processor.

Step 2 Connect the male end of the A&D-supplied RS-232C communication cable to the processor.

Step 3 Connect the female end of the A&D-supplied RS-232C communication cable to the computer.

Step 4 The display on the processor will show “0-----”. If the processor does not have a

display, the LED should be green. If the LED is not green or the “0-----” does not

appear, see 3-2 Communication Errors”. If the processor shows “0-----”, go to step

5. If the processor has a green light, go to step 6.

Note: If possible, keep the cable connected to the computer unless the port is needed for other functions.

Step 5 Press the ENTER key on the processor keypad. A series of 6 dashes “------” will

appear on the processor screen. Step 6 Run the software.

Step 7 Confirm that the correct serial port is set. For details, see 2-9-4-2 Serial Port”.

2-2-2 Connecting the TM-2430 series to the computer

You must use the A&D-supplied RS-232C communication cable, otherwise communication errors will occur.

You must use AX-KO1502 cable for 9 pin serial port (supplied) or AX-KO1503 cable for 25 pin serial port (available separately as an option).

Figure 2 : Connection between the TM-2430 series and the computer

(The illustration above is not to scale.)

Step 1 Connect the male end of the A&D-supplied RS-232C communication cable to the monitor.

Step 2 Connect the female end of the A&D-supplied RS-232C communication cable to the computer.

7

Image 8
Contents TM-2430-13 Contents Troubleshooting Introduction About the Data Analysis SoftwareSymbols Monitoring Device ReviewInstalling the Software SafetySoftware Package Components System RequirementsConnecting the TM-2420/TM-2421 to the computer Getting StartedCommunication Between the Monitor and the Software Communication Between the Monitor and the SoftwareConnecting the TM-2430 series to the computer Connecting the UA-767PC to the computer Main Screen Description Analysis Intervals BP Mode ToolbarMethod Setting Description File Menu Menu DescriptionSave As DeleteSelect the Merge the BP data files sub-menu ConvertFile Merge Print Click OpenExit Saving a FileView Menu Summary Data display BP Blood Errors forSummary Data Tab Definition See 3-1 Measurement Errors StatusExcluding/Including BP Measurements Entering CommentsLinking to Trends Screen Systolic/Diastolic Blood Pressure BP TrendPulse Pulse Trend Mean Arterial Blood Pressure MAP TrendDelimitation, Scroll Bar and Linking Partial Analysis SleepBlood Pressure Trend BP Limits Scroll bar/LinkingTrend Display Positioning Trends Viewing the TrendsScaling the Screen Trend Page Setup dialog box Creating a New/Edit Trends TemplateBlood Pressure Trend Setup dialog box Correlation Plots display Systolic Blood Pressure Diastolic Blood PressureMAP, Mean Blood Arterial Pressure PulseCircadian Rhythm display HIB Analysis HIB analysis displayField Description Intervals and ConditionsData retrieval. See 2-7-1 Retrieving Data from the Monitor Physician ListRecorder Menu Select the Patient Information sub-menuClick the Retrieve Data button Retrieving Data from the MonitorSelect the Recorder menu Select the Retrieve data sub-menu Clearing The Monitor’s Memory Select the Recorder menuMonitoring session is performed at the same interval for TM-2420/TM-2421 Recorder Setup dialog box 3-2 TM-2430 Series Setup Description of each mode is as followsPeriod Start Interval Example Programming the Monitor for Three 3 Periods Monitor programming exampleExample Programming the Monitor for Four 4 Periods OFF3-4 UA-767PC Setup Blood Pressure Measurement Available only with the UA-767PCReport Menu Report dialog boxReport setup dialog box Creating a New Report TemplateSaving a Default Report Template for Printing Editing an Existing Report TemplateDeleting an Existing Report Template Printing an Existing Report TemplateOptions Menu Adding a New Physician to the List Editing an Existing Physician in the ListDeleting an Existing Physician in the List Trend Display See 2-6-3-3 Trend Display Positioning TrendsDefault Blood Pressure Settings dialog box Recorder Type On-line HelpSerial Port Measurement Errors 1 TM-2420/TM-2421 Error CodesBe erased Well as programmed intervalsBe taped to the arm with the supplied 2 TM-2430 Series Error Codes Communication Errors Communications Error dialog boxSummary Data / Sleep Tab Data can not be Seen Reports Are Not Printing With the Desired BP ModeMissing Files Printout Examples Summary data Full and partial Text data Summary data Awake and sleep Text data Blood pressure data Text data Trend graph Blood pressure and double product Graphical data Page Page Mini report Mini report Terms Related to Blood Pressure Terms Related to StatisticsTerms Related to Biological Rhythm Mesor95% confidence interval is expressed Terms Related to the Risk Management of High Blood Pressure Memo Memo Memo