A&D TM-2430-13 instruction manual Be taped to the arm with the supplied

Page 50

ERROR CODE

POSSIBLE CAUSE

 

CORRECTION

 

 

 

 

E07

Manual stop

 

 

The

measurement

was

manually

 

 

 

 

 

interrupted by depression of the red

 

 

 

 

 

Start/Stop button.

 

 

 

E08

Korotkoff

 

sounds

are

Check that the lower microphone is

 

undetectable.

 

 

affixed with the sensing side (skin side)

 

 

 

 

 

facing the patient’s arm.

 

 

 

 

 

 

 

 

Note: The microphone should always

 

 

 

 

 

be taped to the arm with the supplied

 

 

 

 

 

affixing tape. Place "X" of tape over the

 

 

 

 

 

top of the microphone.

 

 

 

E11

More than 256 Korotkoff sounds

Check for a break in the microphone

 

are measured.

 

 

lead. Check for external damage to the

 

 

 

 

 

microphone. Microphone circuit may be

 

 

 

 

 

damaged.

 

 

 

 

 

 

 

 

 

 

Note: If a break in the microphone lead

 

 

 

 

 

or external damage has occurred, the

 

 

 

 

 

microphone

assembly

must

be

 

 

 

 

 

replaced.

 

 

 

 

 

E20

Pulse Error

 

 

Check the position of the lower

 

Pulse less than 35 beats or more

microphone and then re-correlate it to

 

than 200 beats per minute.

 

the subject.

 

 

 

 

 

 

 

 

 

Note: Certain physiological conditions

 

 

 

 

 

or

pharmaceutical treatment

may

 

 

 

 

 

excessively reduce or increase a pulse

 

 

 

 

 

rate. In cases of excessively low pulse

 

 

 

 

 

rates, have the subject walk or stand a

 

 

 

 

 

few minutes to raise the pulse rate

 

 

 

 

 

above 35 beats per minute.

 

 

E21

Diastolic greater than 160 mmHg

Outside

of

normal

physiological

 

or less than 40 mmHg.

 

parameters.

 

 

 

 

E22

Systolic greater than 280 mmHg

Outside

of

normal

physiological

 

or less than 60 mmHg.

 

parameters.

 

 

 

 

E23

Difference between systolic blood

Outside

of

normal

physiological

 

pressure

and

diastolic

blood

parameters.

 

 

 

 

 

pressure less than 10 mmHg.

 

 

 

 

 

 

 

E24

Difference between systolic blood

Confirm that the subject is not moving

 

pressure

and

diastolic

blood

during measurement.

Reposition

the

 

pressure

greater than

150

lower microphone and then correlate it

 

mmHg.

 

 

 

to the subject.

 

 

 

 

Note: If a measurement error occurs (other than E08) when the “Auto” mode is set, the monitor will automatically take a 2nd measurement. If a second measurement error occurs, a 3rd measurement will not be taken. The monitor will return to its programmed interval.

49

Image 50
Contents TM-2430-13 Contents Troubleshooting Introduction About the Data Analysis SoftwareSymbols Monitoring Device ReviewSoftware Package Components Installing the SoftwareSafety System RequirementsCommunication Between the Monitor and the Software Connecting the TM-2420/TM-2421 to the computerGetting Started Communication Between the Monitor and the SoftwareConnecting the TM-2430 series to the computer Connecting the UA-767PC to the computer Main Screen Description Analysis Intervals BP Mode ToolbarMethod Setting Description File Menu Menu DescriptionSave As DeleteSelect the Merge the BP data files sub-menu ConvertFile Merge Exit PrintClick Open Saving a FileView Menu Summary Data display BP Blood Errors forSummary Data Tab Definition See 3-1 Measurement Errors StatusExcluding/Including BP Measurements Entering CommentsPulse Pulse Trend Linking to Trends ScreenSystolic/Diastolic Blood Pressure BP Trend Mean Arterial Blood Pressure MAP TrendBlood Pressure Trend BP Limits Delimitation, Scroll Bar and Linking Partial AnalysisSleep Scroll bar/LinkingTrend Display Positioning Trends Viewing the TrendsScaling the Screen Trend Page Setup dialog box Creating a New/Edit Trends TemplateBlood Pressure Trend Setup dialog box Correlation Plots display MAP, Mean Blood Arterial Pressure Systolic Blood PressureDiastolic Blood Pressure PulseCircadian Rhythm display HIB Analysis HIB analysis displayData retrieval. See 2-7-1 Retrieving Data from the Monitor Field DescriptionIntervals and Conditions Physician ListRecorder Menu Select the Patient Information sub-menuClick the Retrieve Data button Retrieving Data from the MonitorSelect the Recorder menu Select the Retrieve data sub-menu Clearing The Monitor’s Memory Select the Recorder menuMonitoring session is performed at the same interval for TM-2420/TM-2421 Recorder Setup dialog box 3-2 TM-2430 Series Setup Description of each mode is as followsPeriod Start Interval Example Programming the Monitor for Three 3 Periods Monitor programming exampleExample Programming the Monitor for Four 4 Periods OFF3-4 UA-767PC Setup Blood Pressure Measurement Available only with the UA-767PCReport Menu Report dialog boxReport setup dialog box Creating a New Report TemplateDeleting an Existing Report Template Saving a Default Report Template for PrintingEditing an Existing Report Template Printing an Existing Report TemplateOptions Menu Deleting an Existing Physician in the List Adding a New Physician to the ListEditing an Existing Physician in the List Trend Display See 2-6-3-3 Trend Display Positioning Trends Default Blood Pressure Settings dialog box Recorder Type On-line HelpSerial Port Be erased Measurement Errors1 TM-2420/TM-2421 Error Codes Well as programmed intervalsBe taped to the arm with the supplied 2 TM-2430 Series Error Codes Communication Errors Communications Error dialog boxSummary Data / Sleep Tab Data can not be Seen Reports Are Not Printing With the Desired BP ModeMissing Files Printout Examples Summary data Full and partial Text data Summary data Awake and sleep Text data Blood pressure data Text data Trend graph Blood pressure and double product Graphical data Page Page Mini report Mini report Terms Related to Blood Pressure Terms Related to StatisticsTerms Related to Biological Rhythm Mesor95% confidence interval is expressed Terms Related to the Risk Management of High Blood Pressure Memo Memo Memo