ERROR CODE | POSSIBLE CAUSE |
| CORRECTION |
|
|
|
| ||||
E07 | Manual stop |
|
| The | measurement | was | manually | ||||
|
|
|
|
| interrupted by depression of the red | ||||||
|
|
|
|
| Start/Stop button. |
|
|
| |||
E08 | Korotkoff |
| sounds | are | Check that the lower microphone is | ||||||
| undetectable. |
|
| affixed with the sensing side (skin side) | |||||||
|
|
|
|
| facing the patient’s arm. |
|
|
| |||
|
|
|
|
| Note: The microphone should always | ||||||
|
|
|
|
| be taped to the arm with the supplied | ||||||
|
|
|
|
| affixing tape. Place "X" of tape over the | ||||||
|
|
|
|
| top of the microphone. |
|
|
| |||
E11 | More than 256 Korotkoff sounds | Check for a break in the microphone | |||||||||
| are measured. |
|
| lead. Check for external damage to the | |||||||
|
|
|
|
| microphone. Microphone circuit may be | ||||||
|
|
|
|
| damaged. |
|
|
|
|
| |
|
|
|
|
| Note: If a break in the microphone lead | ||||||
|
|
|
|
| or external damage has occurred, the | ||||||
|
|
|
|
| microphone | assembly | must | be | |||
|
|
|
|
| replaced. |
|
|
|
|
| |
E20 | Pulse Error |
|
| Check the position of the lower | |||||||
| Pulse less than 35 beats or more | microphone and then | |||||||||
| than 200 beats per minute. |
| the subject. |
|
|
|
| ||||
|
|
|
|
| Note: Certain physiological conditions | ||||||
|
|
|
|
| or | pharmaceutical treatment | may | ||||
|
|
|
|
| excessively reduce or increase a pulse | ||||||
|
|
|
|
| rate. In cases of excessively low pulse | ||||||
|
|
|
|
| rates, have the subject walk or stand a | ||||||
|
|
|
|
| few minutes to raise the pulse rate | ||||||
|
|
|
|
| above 35 beats per minute. |
|
| ||||
E21 | Diastolic greater than 160 mmHg | Outside | of | normal | physiological | ||||||
| or less than 40 mmHg. |
| parameters. |
|
|
|
| ||||
E22 | Systolic greater than 280 mmHg | Outside | of | normal | physiological | ||||||
| or less than 60 mmHg. |
| parameters. |
|
|
|
| ||||
E23 | Difference between systolic blood | Outside | of | normal | physiological | ||||||
| pressure | and | diastolic | blood | parameters. |
|
|
|
| ||
| pressure less than 10 mmHg. |
|
|
|
|
|
|
| |||
E24 | Difference between systolic blood | Confirm that the subject is not moving | |||||||||
| pressure | and | diastolic | blood | during measurement. | Reposition | the | ||||
| pressure | greater than | 150 | lower microphone and then correlate it | |||||||
| mmHg. |
|
|
| to the subject. |
|
|
|
|
Note: If a measurement error occurs (other than E08) when the “Auto” mode is set, the monitor will automatically take a 2nd measurement. If a second measurement error occurs, a 3rd measurement will not be taken. The monitor will return to its programmed interval.
49