Digi DC-190 installation manual Installation, Locking and Unlocking, Setting Up

Page 7

2.0Installation

This section describes the procedure for the installation and setup of the DC-190 counting scale.

2.1Locking and Unlocking

Do not turn scale upside down. Always Caution work with scale on its side! Damage to the load cell can occur if the scale is

turned upside down.

The DC-190 counting scale is delivered in a locked position to prevent damage to the load cells during shipment.

To prevent damage to the load cells, Caution scale must be locked prior to shipment.

The scale uses either one setscrew for the single-platform scale or two setscrews for the dual-platform scale. The setscrews are located on the bottom of the base and must be removed before the scale is put into service. Use the following procedure to unlock your DC-190 counting scale.

1.Turn scale on side. Loosen locknut 1/4 turn (see Figure 2-1).

SINGLE-PLATFORM LOAD

CELL SETSCREW

DUAL-PLATFORM LOAD

CELL SETSCREWS

Figure 2-1. Location of Load Cell Setscrews for Single- and

Dual-Platform Scales

2.Remove load cell setscrew (see Figure 2-2) using the 2 mm hex wrench provided with scale.

NOTE: Keep locknut in the approximate original position on the setscrew to prevent damage to load cell when reinstalling.

NOTE: Beginning in 2003, setscrews with slotted heads will replace the hex head types, and will no longer require the 2mm hex wrench for removal.

LOAD CELL SETSCREW

Figure 2-2. Setscrew Removal

3.Tape setscrews to the bottom of the scale or store in a safe location for possible future use.

2.2Setting Up

Place the scale on a solid, level surface away from fans, breezes, and sources of electrical interference.

Level the scale by turning the four adjustable legs located on the bottom of the scale while referencing the bubble level located on the back of the scale (see Figure 2-3).

NOTE: To ensure greater scale stability, turn in all four adjustable legs before leveling. Turn out adjustable legs to level as needed.

BUBBLE LEVEL

Figure 2-3. Bubble Level

Image 7
Contents DC-190 Introduction Unpacking and Inspection DC-190 Scale Capacities and ResolutionsRepacking Other S-Series Remote Platforms UW/1000 Pcs DC-190 ConsoleScale no 2 DC-190 Keypad Setting Up InstallationLocking and Unlocking Location of AC Power Cord Receptacle Load Cell Cable Powering Up the DC-190Location of Top Cover Screws Setting Time and DatePole Mounting Instructions Installing Cable Strain ReliefScale Configurations Scale SetupSpec Spec SPEC17 Switching if Spec 2, bit 1 is set Customer Specification 141 SettingsAPW DIN-8 Force Balance 010 Weight and Measurement Specifications 142 Settings Selecting 1 can damage Connector, selectSpec Spec Min MaxMax Spec Min Load Cell Sensitivity Selection mV/V-PORT 2, Scale 2 orNot used set to RS-232 Specifications 143 SettingsScale Operations Entering Tare WeightsEntering Unit Weights Toggling Between Net and Gross WeightToggle Between Scales Unit Weight Operation by Key Entry Other Operations Item Code Storage Scale ProgrammingItem Code Maintenance Using Item Code in Normal Mode Re-Computing Item Code Unit Weight Procedure Pin DIN Female Connector PinoutsDC-190 Messages Clearing Locked-Up Scale 888888 Error 10.11 DC-190 Ultra Count Specifications DC-190 Ultra Count Limited Warranty