4.3Weight and Measurement Specifications (142 Settings)

To configure weight and measurement specifications, press and hold the REZERO key and enter 142 using the numeric keypad. SPC20 appears in WEIGHT display and configuration of four bits (for SPEC 20) appears in QUANTITY display. Follow the procedure described in Section 4.1 on page 17 to update the SPEC settings for the 142 access code.

SPEC

Default

 

Bit 3

 

 

 

Bit 2

 

 

Bit 1

 

Bit 0

 

 

 

 

 

 

 

20

none

Minimum Display (PORT 1, Scale 1)

 

Minimum Display (PORT 1, Scale 2)

 

 

MUST BE SET PRIOR TO CALIBRATION!

 

MUST BE SET PRIOR TO CALIBRATION!

 

 

00:

2

01:

1

 

 

 

00:

2

01:

1

 

 

 

10:

5

11:

10

 

 

 

10:

5

11:

10

 

 

 

 

 

 

 

21

none

Eltron Printer Selection

Weight Decimal Point Position (PORT 1, Scale 1)

 

 

 

0:

Eltron Model 2722

MUST BE SET PRIOR TO CALIBRATION!

 

 

 

 

1:

Eltron Model 2742

000: 00000

011: 00.000

 

 

 

 

 

 

or 2642

 

 

001: 0000.0

100: 0.0000

 

 

 

 

 

NOTE: Must be set to

010: 000.00

101: .00000

 

 

 

 

 

0 if an Eltron printer is

 

 

 

 

 

 

 

 

 

 

not being used.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

22

none

Not used

 

 

Weight Decimal Point Position (PORT 1, Scale 2)

 

 

 

 

 

 

 

MUST BE SET PRIOR TO CALIBRATION!

 

 

 

 

 

 

 

 

000: 00000

011: 00.000

 

 

 

 

 

 

 

 

 

001: 0000.0

100: 0.0000

 

 

 

 

 

 

 

 

 

010: 000.00

101: .00000

 

 

 

 

 

 

 

 

 

 

23

0000

Display Resolution for Scales 1, 2, 3, and 4.

 

Zero Setting Range

 

 

 

(Specify Not Used for force balance.)

 

00: +Unlimited/–10% FS

01: ±2% FS

 

 

 

 

 

 

 

 

 

 

 

00: 1/10000

 

01: 1/5000

 

10: ±10% FS

 

11: Not used

 

 

10: 1/2500

 

11: Not used

 

 

 

 

 

 

 

 

 

 

 

 

 

24

0000

Masked Display at

 

Display at Minus Weight

Zero Lamp Lighting

Low Battery (turn off

 

 

Minus Weight

 

 

0:

Minus display

 

Method

 

 

display when low battery)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0:

Gross

 

 

1:

Masked

 

0:

Gross

 

 

0: No

 

 

1:

Net

 

 

 

 

 

1:

Net

 

 

1: Yes

 

 

 

 

 

 

 

 

 

25

00_0

Scale Starting Method

Internal Resolution

 

Scale Type

 

 

Gross Mode Available

 

 

0:

Automatic

 

Protected by Span

 

0:

Single scale

 

0: Yes (allows gross

 

 

 

Switch

 

 

 

 

1:

Manual (must

 

 

1:

Dual scale

 

mode selection from

 

 

 

 

 

 

 

 

 

 

press REZERO key

0:

No (internal

 

NOTE: If you are not

keypad)

 

 

 

on startup)

 

 

resolution can be

1: No (inhibits gross

 

 

 

 

 

connected to a

 

 

 

 

 

 

 

 

viewed by pressing

 

mode; switching not

 

 

 

 

 

 

 

dual-platform scale, or

 

 

 

 

 

 

 

REZERO * * +)

 

allowed)

 

 

 

 

 

 

 

 

do not have two scales

 

 

 

 

 

 

1:

Yes (protected until

 

 

 

 

 

 

 

wired into Cable PORT

 

 

 

 

 

 

 

 

internal span

 

 

 

 

 

 

 

 

 

 

1, set this bit to 0 (single

 

 

 

 

 

 

 

 

switch pressed)

 

 

 

 

 

 

 

 

 

 

scale). Selecting 1 (dual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

scale) can damage the

 

 

 

 

 

 

 

 

 

 

console’s electronic

 

 

 

 

 

 

 

 

 

 

components.

 

 

 

 

 

 

 

 

 

 

 

26

0000

Zero Tracking When

Rezero with a Tare

 

Initial Start Range

 

 

 

 

Tare Is Present

 

Weight

 

00: +Unlimited/–10% FS

01: ±2% FS

 

 

 

 

 

 

 

 

 

 

 

0: Yes

 

 

0: Yes

 

10: ±10% FS

 

11: Not used

 

 

1: No (zero tracking is

1: No

 

 

 

 

 

 

 

 

off with tare in system)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 4-2. DC-190 142 Settings

Image 16
Contents DC-190 Introduction Unpacking and Inspection DC-190 Scale Capacities and ResolutionsRepacking Other S-Series Remote Platforms UW/1000 Pcs DC-190 ConsoleScale no 2 DC-190 Keypad Setting Up InstallationLocking and Unlocking Powering Up the DC-190 Location of AC Power Cord Receptacle Load Cell CableSetting Time and Date Location of Top Cover ScrewsInstalling Cable Strain Relief Pole Mounting InstructionsScale Configurations Scale SetupSpec Spec SPEC17 Switching if Spec 2, bit 1 is set Customer Specification 141 SettingsAPW DIN-8 Force Balance 010 Weight and Measurement Specifications 142 Settings Connector, select SpecSpec Min Max Selecting 1 can damageLoad Cell Sensitivity Selection mV/V-PORT 2, Scale 2 or Max Spec MinRS-232 Specifications 143 Settings Not used set toEntering Tare Weights Scale OperationsEntering Unit Weights Toggling Between Net and Gross WeightToggle Between Scales Unit Weight Operation by Key Entry Other Operations Scale Programming Item Code StorageItem Code Maintenance Using Item Code in Normal Mode Re-Computing Item Code Unit Weight Procedure Connector Pinouts Pin DIN FemaleDC-190 Messages Clearing Locked-Up Scale 888888 Error 10.11 DC-190 Ultra Count Specifications DC-190 Ultra Count Limited Warranty