3.0Scale Setup

This section provides information about attaching scales and serial devices to the DC-190. Information presented describes both physical connections and values that must be specified when configuring the DC-190 (see Section 4.0). If you know what scales and serial devices will be connected to the DC-190, you can use the SPEC Code Worksheets Section 10.1 on page 50 to record these values for later configuration.

3.1Scale Configurations

The DC-190 can be connected to up to three analog scale bases using the load cell ports (PORT 1 and PORT 2 shown in Figure 3-1). An Ohaus® Explorer force balance can also connected to either the FORCE BALANCE or 232/PRINTER serial port for a total of four scale channels.

PORT 1

PORT 2

Figure 3-1. DC-190 Scale Ports and Connectors

Attach SX-L single and SX-D dual scale platforms to PORT 1 (left connector); attach external scale base to PORT 2 (right connector). For three-scale configurations using either an SX-L and two external bases or three external bases, an optional Y-cable must be used to connect two of the bases to PORT 1. (A wiring diagram of the Y-cable is shown in Section 10.5 on page 55.)

3.1.1Configuring SPECs 16 and 17

SPECs 16 and 17 are used to assign the annunciators for each of the four scale channels (Scale No. annunciator on the DC-190 console). All scale channels must be assigned, regardless of the actual number of scales attached. Table 3-1 shows the values and default bit configurations for SPECs 16 and 17.

SPEC

Default Setting

 

Bits 3 and 2

 

Bits 1 and 0

 

 

 

 

16

0001

Scale 1

Scale 2

 

 

00:

Internal Scale 1

00:

Internal Scale 1

 

 

01:

Internal Scale 2

01:

Internal Scale 2

 

 

10:

External Scale

10:

External Scale

 

 

11:

Force Balance

11:

Force Balance

 

 

 

 

17

1011

Scale 3

Scale 4

 

 

00:

Internal Scale 1

00:

Internal Scale 1

 

 

01:

Internal Scale 2

01:

Internal Scale 2

 

 

10:

External Scale

10:

External Scale

 

 

11:

Force Balance

11:

Force Balance

 

 

 

 

 

 

Table 3-1. Specifications 16 and 17 Default Settings

When configuring SPECs 16 and 17, use the following rules:

If a force balance is attached, it must be configured as Scale 1. That is, SPEC 16, bits 3 and 2, (Scale 1) must be set to Force Balance (11).

Scales attached to Port 1 are configured as Internal Scale 1 and Internal Scale 2, regardless of whether the actual scale base is internal (SX-L, SX-D) or external.

Image 11
Contents DC-190 Introduction Repacking DC-190 Scale Capacities and ResolutionsUnpacking and Inspection Other S-Series Remote Platforms Scale no DC-190 ConsoleUW/1000 Pcs 2 DC-190 Keypad Locking and Unlocking InstallationSetting Up Location of AC Power Cord Receptacle Load Cell Cable Powering Up the DC-190Location of Top Cover Screws Setting Time and DatePole Mounting Instructions Installing Cable Strain ReliefSpec Scale SetupScale Configurations Spec SPEC17 APW Customer Specification 141 SettingsSwitching if Spec 2, bit 1 is set DIN-8 Force Balance 010 Weight and Measurement Specifications 142 Settings Selecting 1 can damage Connector, selectSpec Spec Min MaxMax Spec Min Load Cell Sensitivity Selection mV/V-PORT 2, Scale 2 orNot used set to RS-232 Specifications 143 SettingsScale Operations Entering Tare WeightsToggle Between Scales Toggling Between Net and Gross WeightEntering Unit Weights Unit Weight Operation by Key Entry Other Operations Item Code Storage Scale ProgrammingItem Code Maintenance Using Item Code in Normal Mode Re-Computing Item Code Unit Weight Procedure Pin DIN Female Connector PinoutsDC-190 Messages Clearing Locked-Up Scale 888888 Error 10.11 DC-190 Ultra Count Specifications DC-190 Ultra Count Limited Warranty