A&D 1WMPD4002313 manual Function list

Page 13

6-2. Function list

Item

Function number

 

Description

 

 

Š C01- 0

CAL switch enabled,

0

key enabled

Do not set

Calibration

C01- 1

CAL switch disabled,

0

key disabled

“C01-1”

 

C01- 2

CAL switch enabled,

0

key disabled

 

 

 

Š C02- 0

6 kg (15 lb)

 

 

 

 

C02- 1

15 kg (35 lb)

 

 

 

 

C02- 2

30 kg (70 lb)

 

 

 

Capacity

C02- 3

60 kg (150 lb)

 

 

 

C02- 4

150 kg (350 lb)

 

 

 

 

 

 

 

 

C02- 5

300 kg (700 lb)

 

 

 

 

C02- 6

600 kg (1500 lb)

 

 

 

 

C02- 7

1500 kg (3500 lb)

 

 

 

Weighing display

C03- 0

kg, g

 

 

See 11-1.

unit

Š C03- 1

kg, g, lb, oz, lb-oz

 

 

Capacity ...”.

Power-on-zero

C04- 0

±10% of the capacity

 

 

OIML R76 / NTEP

range

Š C04- 1

±50% of the capacity

 

 

 

Internal use

Š C05- 0

Do not change.

 

 

 

Decimal

Š C06- 0

. (point)

 

 

 

separator

C06- 1

, (comma)

 

 

 

Tare clear by

Š C07- 0

ZERO does not clear tare weight

 

ZERO operation

C07- 1

ZERO clears tare weight

 

OIML R76

Tare operation

C08- 0

Tare weight is rounded to the weight

NTEP

(Net weight

display count

 

 

 

 

 

 

calculation)

Š C08- 1

Internal count is used for TARE weight.

OIML R76

Preset tare

C09- 0

Function disabled

 

 

 

function

Š C09- 1

Function enabled

 

 

 

lb-oz unit

C10- 0

lb-oz disabled

 

 

 

Š C10- 1

lb-oz enabled

 

 

 

 

 

 

 

ZERO operation

Š C11- 0

±2% of the capacity

 

 

 

range

C11- 1

Up to the capacity

 

 

 

 

Š C12- 0

“f02” setting specifies the resolution.

 

Display resolution

C12- 1

Fixed to “f02-0” (1/3,000~1/3,500)

 

C12- 2

Fixed to “f02-1” (1/6,000~1/7,500)

 

 

 

 

C12- 3

Fixed to “f02-2” (1/12,000~1/15,000)

 

Auto-tare function

C13- 0

Function disabled

 

 

C13-0 deletes

Š C13- 1

Function enabled

 

 

“f21” ~ “f23”.

ŠFactory setting

11

Image 13
Contents Weighing Indicator This Manual and Marks Contents Compliance with FCC rules ComplianceClassification of protection provided by enclosures Introduction Unpacking Load Cell Connections Precautions for Installing the IndicatorCalculating formula Verifying Load Cell Output and Input SensitivityExample of calculation Parameter Settings and Calibration Setting Up the IndicatorNames and Functions Names Functions / Operation Comparison Indicators Procedure for setting parameters Function Parameter Settings00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00 Function list Calibration Calibration using a weight00 Cal00 00Cal0000end00 Go to of 7-1. Calibration using a weightGravity acceleration correction AUTO-TARE Function Additional Information Circuit diagramOP-03 RS-232C 2-3. Command Mode Data output mode f06Data format OP-04 RS-422 Multi-connection with Print key mode new functionLb1 Error MessagesErr-# Specifications SpecificationsCapacity and minimum display Options DimensionsPage German Sales Office UK OfficeKorea Limited Instruments India Private Limited