A&D 1WMPD4002313 manual Function Parameter Settings, Procedure for setting parameters

Page 12

6. FUNCTION PARAMETER SETTINGS

To design a platform scale, set the capacity, weighing unit and other functions according to the specifications of the scale.

Normally, end users do not have to set these function parameters and these setting can be sealed together with the calibration switch.

Parameter

-C01-00

Function number

6-1. The procedure for setting parameters

‰Do not set “C01-1”. Once set, the indicator cannot get into the parameter

setting procedure again.

‰ When turning the power on and the indicator may display ------- ,

press the PRINT/ENT key to proceed to the weighing mode.

‰See “7. CALIBRATION” as for the location of the calibration (CAL) switch.

1.

 

In the weighing mode, press and hold the calibration (CAL)

 

 

switch until

 

Cal

 

 

appears. Then, release the switch.

‰ Press and hold the

 

 

 

key for about 5 seconds, then you

 

0

 

 

can also enter the calibration mode (when “C01-0” is set).

‰ Press the

 

ON/OFF

 

key or (CAL) switch to stop the

 

setting procedure.

 

 

 

 

2.

Press and hold the

 

 

key and press the

 

ZERO

 

 

 

key. Then the first function number is

 

 

PRINT/ENT

 

 

displayed.

 

 

 

 

 

 

 

 

3.Press the 2 key to enter into the “C02 Capacity” setting.

4.Press the PRINT/ENT key. Then the parameter stored appears.

5.Enter the parameter value for the capacity using the 10-key pad.

6.Press the ZERO key to go to the next function number “C03 Weighing display unit”.

7.Set C03 “0” (metric unit), or “1” (non-metric unit).

8.The settings C02 and C03 are minimum requirements to design a platform scale. To finalize the settings, be

sure to press the PRINT/ENTER key. Then, the display shows end for a few seconds and returns to weighing mode.

-00Cal00

-C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00

‰ The ZERO key increments the function number as in the step 6 above, but note that all of the new parameters are not stored until end is displayed.

‰ To exit with nothing changed, turn the power off without pressing the PRINT/ENT key.

‰Set the other parameters according to the specifications required.

10

Image 12
Contents Weighing Indicator This Manual and Marks Contents Compliance Compliance with FCC rulesClassification of protection provided by enclosures Introduction Unpacking Precautions for Installing the Indicator Load Cell ConnectionsVerifying Load Cell Output and Input Sensitivity Calculating formulaExample of calculation Setting Up the Indicator Parameter Settings and CalibrationNames and Functions Names Functions / Operation Comparison Indicators Function Parameter Settings Procedure for setting parameters00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00 Function list Calibration using a weight Calibration00 Cal00 00Cal00Go to of 7-1. Calibration using a weight 00end00Gravity acceleration correction AUTO-TARE Function Circuit diagram Additional InformationOP-03 RS-232C Data output mode f06 2-3. Command ModeData format Multi-connection with Print key mode new function OP-04 RS-422Error Messages Lb1Err-# Specifications SpecificationsCapacity and minimum display Dimensions OptionsPage UK Office German Sales OfficeKorea Limited Instruments India Private Limited