8. AUTO-TARE FUNCTION

The FS-D has an auto-tare function to be used with the comparator function enabled. If the weight value is within the OK range and stable for a preset time, then the indicator will automatically tare the weight and show zero. The Function “f22” designates the timing to tare automatically. See “9. ADDITIONAL INFORMATION” about the added F-Function settings related to the auto-tare function.

‰To use the auto-tare function, set the function settings below.

f08-1: Compare all weighing data (other settings may be used depending on the application).

f21-1: Auto-tare function enabled.

f22-0 ~ 9: Select the timing to tare automatically to avoid the wrong tare operation, for example; too early to tare, to take a longer time to go to the next weighing and so on.

‰Take-away check weighing:

Take-away check weighing (negative comparison) is the way to compare the negative weight taking away material from a container. Set the Function “f24-1” together with the auto-tare function enabled “f21-1”. In this operation mode, the indicator operates as “take-away the material” Æ “OK and stable” Æ “auto-tare” Æ “take-away the material” Æ ······.

In this setting, the polarity of the target weight and the upper / lower limit weight are ignored and the indicator shows the comparator results as below.

f07-1:

-(Target+HI limit)

-(Target-LO limit)

f07-0:

-Upper limit

-Lower limit

Negative value

Net 0 Weight display

HI

OK

LO

Note To start the take-away check weighing, be sure to use the

 

TARE

key to tare

the weight of the container filled with material. The

 

 

key may zero the

ZERO

 

display, and the indicator goes below the zero point by taking out the material. Then, the auto-tare function does not work.

‰When the Function “f23-1 Tares the initial (container) weight.” is set:

To start the auto-tare function application, usually the container (filled with material) will be placed on the weighing pan and its weight must be tared using the TARE key. When the Function “f23-1” is set, the indicator will tare the initial (container) weight automatically.

Remove all load on the weighing pan to return to the zero point, the tare weight will be automatically cleared. If the indicator could not return to the zero point because of a

zero shift, press the ZERO key to clear the tare weight.

‰If the indicator is equipped with the optional RS-232C interface, the OK weighing data only can be sent out automatically. Set the Function setting “f06-7 Auto-print mode +/- data”.

14

Image 16
Contents Weighing Indicator This Manual and Marks Contents Compliance with FCC rules ComplianceClassification of protection provided by enclosures Introduction Unpacking Precautions for Installing the Indicator Load Cell ConnectionsCalculating formula Verifying Load Cell Output and Input SensitivityExample of calculation Setting Up the Indicator Parameter Settings and CalibrationNames and Functions Names Functions / Operation Comparison Indicators Procedure for setting parameters Function Parameter Settings00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00 Function list Calibration using a weight Calibration00 Cal00 00Cal0000end00 Go to of 7-1. Calibration using a weightGravity acceleration correction AUTO-TARE Function Additional Information Circuit diagramOP-03 RS-232C 2-3. Command Mode Data output mode f06Data format Multi-connection with Print key mode new function OP-04 RS-422Lb1 Error MessagesErr-# Specifications SpecificationsCapacity and minimum display Dimensions OptionsPage UK Office German Sales OfficeKorea Limited Instruments India Private Limited