7. CALIBRATION
This function adjusts the scale (indicator) for accurate weighing.
Calibration must be done when the
CAL switch cover
When the scale (indicator) has been moved.
When the ambient environment has significantly changed.
For regular calibration.
Loose the lock screws on the rear side of the display pod, and remove the (CAL) switch cover. The calibration switch inside.
Calibration (CAL) switch
7-1. Calibration using a weight
1.Warm up the indicator with a weighing platform connected for at least half an hour. Change the user function setting “f01” or place something on the platform to
disable the auto
When turning the power on and the indicator may display | , | ||||||||||
| press the |
|
|
|
|
| key to proceed to the weighing mode. | ||||
|
| PRINT/ENT | |||||||||
2. Press and hold the calibration (CAL) switch until |
| ||||||||||
| Cal | appears, and release the switch. | |||||||||
| Press and hold the |
|
|
| key for about 5 seconds, then you | ||||||
| 0 | ||||||||||
| can also enter the calibration mode (when |
|
| ||||||||
Press the |
| ON/OFF |
| key or (CAL) switch to exit without |
|
| |||||
| calibration. |
|
|
|
|
|
|
|
3. Press the | PRINT/ENT | key, then the gravity |
acceleration value will be displayed.
It is not necessary to set the gravity acceleration value when calibrating the indicator with a calibration weight at the place where it is to be used. (See the next section about the gravity acceleration correction.)
4.Press the PRINT/ENT key, then Cal 0 will be displayed.
5.Make sure that there is nothing on the platform, and wait until the STABLE indicator turns on.
6.Press the PRINT/ENT key. The indicator inputs the
zero point, and the display shows 5pn 1 and the weight value for calibration (SPAN calibration).
The weight value is equal to the capacity. When you enter with “kg” or “g” mode, then the value is “kg”. Entering with “lb” or “oz”, then “lb”.
-00Cal00
-05pn010 -0003000.
12