A&D 1WMPD4002313 manual 00 Cal00, 00Cal00, Calibration using a weight

Page 14

7. CALIBRATION

This function adjusts the scale (indicator) for accurate weighing.

Calibration must be done when the FS-D is initially connected to a load cell (base unit). Calibration may also be required according the changes in the environment.

CAL switch cover

‰When the scale (indicator) has been moved.

‰When the ambient environment has significantly changed.

‰ For regular calibration.

Loose the lock screws on the rear side of the display pod, and remove the (CAL) switch cover. The calibration switch inside.

Calibration (CAL) switch

7-1. Calibration using a weight

1.Warm up the indicator with a weighing platform connected for at least half an hour. ‰ Change the user function setting “f01” or place something on the platform to

disable the auto power-off function.

‰ When turning the power on and the indicator may display

-------

,

 

press the

 

 

 

 

 

key to proceed to the weighing mode.

 

 

PRINT/ENT

2. Press and hold the calibration (CAL) switch until

-00Cal00

 

Cal

appears, and release the switch.

‰

Press and hold the

 

 

 

key for about 5 seconds, then you

 

0

 

can also enter the calibration mode (when “C01-0” is set).

 

 

‰ Press the

 

ON/OFF

 

key or (CAL) switch to exit without

 

 

 

calibration.

 

 

 

 

 

 

 

3. Press the

PRINT/ENT

key, then the gravity

acceleration value will be displayed.

‰It is not necessary to set the gravity acceleration value when calibrating the indicator with a calibration weight at the place where it is to be used. (See the next section about the gravity acceleration correction.)

4.Press the PRINT/ENT key, then Cal 0 will be displayed.

5.Make sure that there is nothing on the platform, and wait until the STABLE indicator turns on.

6.Press the PRINT/ENT key. The indicator inputs the

zero point, and the display shows 5pn 1 and the weight value for calibration (SPAN calibration).

‰The weight value is equal to the capacity. When you enter with “kg” or “g” mode, then the value is “kg”. Entering with “lb” or “oz”, then “lb”.

-0097985.

-00Cal00

-05pn010 -0003000.

12

Image 14
Contents Weighing Indicator This Manual and Marks Contents Classification of protection provided by enclosures ComplianceCompliance with FCC rules Introduction Unpacking Precautions for Installing the Indicator Load Cell ConnectionsExample of calculation Verifying Load Cell Output and Input SensitivityCalculating formula Setting Up the Indicator Parameter Settings and CalibrationNames and Functions Names Functions / Operation Comparison Indicators 00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00 Function Parameter SettingsProcedure for setting parameters Function list 00 Cal00 Calibration using a weightCalibration 00Cal00Gravity acceleration correction Go to of 7-1. Calibration using a weight00end00 AUTO-TARE Function OP-03 RS-232C Circuit diagramAdditional Information Data format Data output mode f062-3. Command Mode Multi-connection with Print key mode new function OP-04 RS-422Err-# Error MessagesLb1 Capacity and minimum display SpecificationsSpecifications Dimensions OptionsPage Korea Limited UK OfficeGerman Sales Office Instruments India Private Limited