A&D 1WMPD4002313 manual Specifications, Capacity and minimum display

Page 21

11. SPECIFICATIONS

11-1. Capacity and minimum display

 

C02- setting

0*

1

2

3

4

5

6

7

 

 

Capacity

6

15

30

60

150

300

600

1500

kg

 

 

0.002*

0.005

0.01

0.02

0.05

0.1

0.2

0.5

 

Min. display

0.001

0.002

0.005

0.01

0.02

0.05

0.1

0.2

 

 

 

 

 

0.0005

0.001

0.002

0.005

0.01

0.02

0.05

0.1

 

 

Capacity

6000

15000

30000

 

 

 

 

 

g

 

 

2*

5

10

NA

NA

NA

NA

NA

 

Min. display

1

2

5

 

 

 

 

 

 

 

 

 

 

0.5

1

2

 

 

 

 

 

 

 

Capacity

15

35

70

150

350

700

1500

3500

lb

 

 

0.005*

0.01

0.02

0.05

0.1

0.2

0.5

1

 

Min. display

0.002

0.005

0.01

0.02

0.05

0.1

0.2

0.5

 

 

 

 

 

0.001

0.002

0.005

0.01

0.02

0.05

0.1

0.2

 

 

Capacity

240

560

1120

2400

5600

11200

 

 

oz

 

 

0.1*

0.2

0.5

1

2

5

NA

NA

 

Min. display

0.05

0.1

0.2

0.5

1

2

 

 

 

 

 

 

 

0.02

0.05

0.1

0.2

0.5

1

 

 

lb-

 

Capacity

15

35

70

NA

NA

NA

NA

NA

oz

 

Min. display

0.1

0.1

0.1

 

 

 

 

 

 

 

 

 

*: Factory setting

 

NA: Not Available

 

11-2. Specifications

Input sensitivity

 

0.2 µV / d (d: the smallest Min. display of kg)

Input signal range

0 mV ~ 13 mV

Load cell excitation

5 V DC ±5% (no remote sensing)

voltage

 

Load cell drive capacity

Up to 4 x 350 ohm load cells

Temperature

 

Zero

±(0.2 µV ±0.001% of zero adjustment voltage) / °C (typ.)

coefficient

 

Span

±0.001% / °C of reading (typ.)

Non-linearity

 

±0.01 % of full scale

Display

 

7 segment LCD display (Character height 18.6 mm) with backlight

 

60 segment analog sweep display with backlight

 

 

 

Display update

 

Approximately 20 times per second

Operating temp.

-10°C ~ 40°C / 14°F ~ 104°F, less than 85% R.H.

Power supply

 

AC main (100V ~ 240 V, 50/60 Hz) or SLA Battery (option)

Dimension

 

250 (W) x 202 (D) x 266 (H) mm / 9.8 (W) x 8.0 (D) x 10.5 (H) in.

Weight

 

Approximately 2.7 kg

 

 

 

19

Image 21
Contents Weighing Indicator This Manual and Marks Contents Compliance Compliance with FCC rulesClassification of protection provided by enclosures Introduction Unpacking Load Cell Connections Precautions for Installing the IndicatorVerifying Load Cell Output and Input Sensitivity Calculating formulaExample of calculation Parameter Settings and Calibration Setting Up the IndicatorNames and Functions Names Functions / Operation Comparison Indicators Function Parameter Settings Procedure for setting parameters00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00 Function list Calibration Calibration using a weight00 Cal00 00Cal00Go to of 7-1. Calibration using a weight 00end00Gravity acceleration correction AUTO-TARE Function Circuit diagram Additional InformationOP-03 RS-232C Data output mode f06 2-3. Command ModeData format OP-04 RS-422 Multi-connection with Print key mode new functionError Messages Lb1Err-# Specifications SpecificationsCapacity and minimum display Options DimensionsPage German Sales Office UK OfficeKorea Limited Instruments India Private Limited