A&D 1WMPD4002313 manual 00end00, Gravity acceleration correction

Page 15

‰If you need ZERO calibration but do not need SPAN calibration, turn the power off to exit from the calibration procedure.

7.To calibrate with a different weight, change the displayed value using the 10-key pad.

‰If you enter the wrong number, press the C key. The value returns to the capacity and enter the number again.

8.Place a calibration weight on the platform with the same value as displayed, and wait until the STABLE indicator turns on.

9.Press the PRINT/ENT key. The indicator calibrates

SPAN and end will appear. Remove the weight from the platform, and turn the power off.

-0002000.

-00end00

‰Note

The value set in step 7 will be cleared after the power is switched off.

If the indicator will be moved to another location, set the gravity acceleration value for the current location and calibrate the indicator according to the procedure above. See the next section to set the value.

7-2. Gravity acceleration correction

When the indicator will be moved to different place from the calibration location, it should be calibrated using a calibration weight.

But if a calibration weight is not available, the gravity acceleration correction will compensate the indicator. To use this function, set the gravity acceleration value to the value of the location where the calibration will be done. Refer to the gravity acceleration map appended to the instruction manual of the FS-iseries.

1.At step 3 of the previous section 7-1. Calibration using a weight”, enter a new value using 10-key pad.

‰The integral part “9” is fixed, enter the value after the decimal point.

‰If you hit the wrong number, press the C key. The value returns to the original number. Enter the number again.

2.Press the PRINT/ENT key, then Cal 0 will be displayed.

3.Go to step 5 of 7-1. Calibration using a weight”.

-0098050.

-00Cal00

13

Image 15
Contents Weighing Indicator This Manual and Marks Contents Compliance Compliance with FCC rulesClassification of protection provided by enclosures Introduction Unpacking Load Cell Connections Precautions for Installing the IndicatorVerifying Load Cell Output and Input Sensitivity Calculating formulaExample of calculation Parameter Settings and Calibration Setting Up the IndicatorNames and Functions Names Functions / Operation Comparison Indicators Function Parameter Settings Procedure for setting parameters00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00 Function list 00Cal00 Calibration using a weightCalibration 00 Cal00Go to of 7-1. Calibration using a weight 00end00Gravity acceleration correction AUTO-TARE Function Circuit diagram Additional InformationOP-03 RS-232C Data output mode f06 2-3. Command ModeData format OP-04 RS-422 Multi-connection with Print key mode new functionError Messages Lb1Err-# Specifications SpecificationsCapacity and minimum display Options DimensionsPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited