The frame ground and the shield are connected internally and there is no difference between them in use.

Data format

‰Example of data

Out of range “kg” (+)

O L

,

+ 9 9 9 9 9 9 9 9 _ k g CR

LF

Out of range “kg” (+)

O L

,

+ 9 9 9 9

.

9 9 9 _ k g CR

LF

The position of decimal point is different according to the model and/or “f02”.

Example for the weighing unit “lb-oz” (added)

Weighing data “lb-oz” (+)

S T

,

+ 0 0 1 L 0 1

.

6 _ o z CR

LF

Data output mode (f06)

The commands can be used in all data output modes (f06-1 is command mode only). The new modes f06-5, f06-6 and f06-7 are added to the Function “f06”.

‰Auto-print Mode + data and OK (f06-6)

Data is sent if the weight display is stable at +5d (d = weighing display division) and above, and the comparator result OK in addition. The next transmission cannot occur until after the weight display falls below +5d.

‰Auto-print Mode +/- data and OK (f06-7)

Data is sent if the weight display is stable at ±5d (d = weighing display division) and above/below, and the comparator result OK in addition. The next transmission cannot occur until after the weight display falls between 5d and +5d.

‰Multi-connection with Print key mode (RS-422/485) (f06-5)

This is one of the applications when a PC and more than one FS-D (RS-422/485 installed) are connected. Refer to “Multi-connection with Print key mode” in detail.

(In 2-3. Command Mode”)

Examples of command and reply (“_” shows “Space” (20H).) There are corrections to the instruction manual 1WMPD4001368.

‰Set a HI limit value or upper limit weight.

When f07-0 or f07-1 is set, the command should have +” and 6 digit number without decimal point.

‰Set a LO limit value or lower limit weight.

When f07-0 or f07-1 is set, the command should have +” and 6 digit number without decimal point.

(No reply for f20-1.)

When f07-0 or f07-1 is set, the command should have +/- (“+” only for f07-1) and 6 digit number without decimal point.

(No reply for f20-1.)

When f07-0 or f07-1 is set, the command should have +/- (“+” only for f07-1) and 6 digit number without decimal point.

When f07-2 is set, the command should have “+” and 5 digit number without decimal point.

16

Image 18
Contents Weighing Indicator This Manual and Marks Contents Compliance Compliance with FCC rulesClassification of protection provided by enclosures Introduction Unpacking Precautions for Installing the Indicator Load Cell ConnectionsVerifying Load Cell Output and Input Sensitivity Calculating formulaExample of calculation Setting Up the Indicator Parameter Settings and CalibrationNames and Functions Names Functions / Operation Comparison Indicators Function Parameter Settings Procedure for setting parameters00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00 Function list 00 Cal00 Calibration using a weightCalibration 00Cal00Go to of 7-1. Calibration using a weight 00end00Gravity acceleration correction AUTO-TARE Function Circuit diagram Additional InformationOP-03 RS-232C Data output mode f06 2-3. Command ModeData format Multi-connection with Print key mode new function OP-04 RS-422Error Messages Lb1Err-# Specifications SpecificationsCapacity and minimum display Dimensions OptionsPage Korea Limited UK OfficeGerman Sales Office Instruments India Private Limited