A&D SW-150KM, SW-6KS, SW-30KM, SW-60KL, SW-60KM, SW-15KM, SW-15KS manual Charging the battery, On/Off

Page 28

1.Disconnect the AC power cable from the electrical outlet.

2.Loosen the M4 screws and remove the rear cover of the display.

Note: Take care not to drop the display.

3.Connect the wires inside the display to the battery with much care so that nothing touches the switches.

Note: Be sure to connect the RED wire to the positive (+ / RED) terminal and the BLACK wire to the negative (- / BLACK) terminal. Or there is a risk of explosion.

4.Install the battery into the display.

5.Attach the rear cover to the display and secure it with the screws loosened at step 2.

6.Connect the AC power cable to the electrical outlet.

7. Press the

ON/OFF

switch and check that the scale turns ON.

8. Disconnect the AC power cable and check that the scale works normally.

13.1.2. Charging the battery

‰When the weight display shows “lb” (Low battery), the battery voltage is low and should be recharged. Turn the scale OFF and connect the AC power cable to an electrical outlet. The charging process will start.

‰Charging will be performed when the AC power cable is connected to an electrical outlet and the scale is turned OFF. If the scale is turned ON, trickle charging will be performed.

‰The scale can be used while the battery is charging. After fully charged, the scale will change the charging process to trickle charge automatically.

Notes

Charge the battery at a temperature between 0°C (32°F) and 40°C (104°F), preferably, at a range of 5°C (41°F) to 35°C (95°F) .

Charge the battery when using for the first time.

The battery must be recharged regularly, every 3 to 6 months, if the scale is not used for a long period of time. More frequent recharging is required in a warmer area.

26

Image 28
Contents SW Series Page Contents Maintenance SpecificationsOptions Gravity Acceleration MAPCompliance with FCC rules ComplianceClassification of protection provided by enclosures Features IntroductionPrecautions Installing the scaleCleaning the scale Operating the scaleUnpacking Description of Each Part Display rear sideDisplay and symbols SymbolsSymbol Description DisplaySwitches Attaching the display pole to the base Setting UPSW-M / L series with a display pole Installing the scale Turning the power on and OFF Basic OperationBat oo ON/OFFIndicator will not turn on Selecting a weighing unitBasic weighing operation During the Tare operationWeight display resolution Counting Mode PrintFive-level comparator mode ComparatorFormula to compare LoloThree-level comparator mode Entering the comparator values Five-level comparator mode Three-level comparator modePrint Sample Tare Calibration Gravity acceleration correctionCalibration using a weight Units PrintRestoring the factory set values Functions Setting the parametersTare ON/OFF Function list Key Installing the battery Using the OP-02 SLA batteryOptions Charging the battery Installing OP-03 OP-03 RS-232C / Relay OutputOP-03 Specifications AsciiSolid sate relay ‰ Data example Command mode Stable positive data Page Installing OP-04 OP-04 RS-422OP-04 Specifications Example of connection RS-422 Stable data Maintenance Error codesSpecifications SpecificationsModel SW-6KS SW-15KS SW-15KM SW-30KM Model SW-60KM SW-60KLExternal dimensions SW-60KL / SW-150KL with a display pole Gravity Acceleration MAP Values of gravity at various locationsWorld map