A&D SW-15KS, SW-6KS, SW-30KM, SW-60KL, SW-60KM, SW-150KM manual Calibration using a weight, Units Print

Page 22

11.2. Calibration using a weight

Prepare a weight, preferably a weight with the same value as the weighing capacity of the scale to be calibrated. Note that the calibration weight value can be changed.

1. Turn the power ON and warm up the scale for at least half an hour.

‰Change the function setting “poff” or place something on the weighing pan to disable the auto power-off function.

‰ Make sure that the scale is in the weighing

Press and hold

 

 

 

 

 

 

 

 

 

 

 

 

mode (displaying “kg” (“g”), “lb”, “oz” or “pcs”).

(more than 4 seconds)

 

 

 

 

 

 

2. Press and hold the

 

 

 

switch until

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TARE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the gravity acceleration

value

appears, and

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

release the switch.

 

 

 

 

Release

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

switch and display “Cal”.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3. Press the

 

UNITS

 

 

 

 

 

Press

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

switch until “Cal 0” is

 

 

 

 

 

4. Press the

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TARE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Confirm that

 

nothing

is placed on the weighing pan and

 

 

 

 

 

 

 

 

 

 

 

 

 

wait for the STABLE indicator to turn on.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5. Press the

 

 

 

switch.

The scale

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TARE

 

Press

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

calibrates the zero point and displays the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

value of the calibration weight (SPAN

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

calibration).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

‰ The calibration weight value is equal to the weighing

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

capacity. (factory setting)

Press

 

 

 

 

 

 

 

 

 

 

 

 

‰ When you enter with “kg” (“g”) or “pcs” mode, the value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

is in “kg”. With “Ib” or “oz”, then “Ib”.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

‰ If SPAN calibration is not to be performed, turn the power OFF to exit from the calibration procedure.

6. To calibrate with a weight different from the weighing

Press

capacity, change the displayed value using the following

 

switches.

 

UNITS

PRINT

To shift the digit that is blinking to the

right.

Calibration

To increase the value of the

weight

 

 

blinking digit by one.

 

 

Weighing pan

 

7.

Place the calibration weight with the same value as

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed on the weighing pan, and wait for the STABLE

 

 

 

 

 

 

 

 

 

indicator to turn on.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8.

Press the

 

switch. The scale calibrates SPAN

 

 

 

 

 

 

 

 

TARE

Press

 

 

 

and “end” is displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Remove the weight from the weighing pan, and turn the

 

 

 

 

 

 

 

 

 

power OFF.

 

 

 

 

 

 

 

 

Note: If the scale will be moved to another location, set the gravity acceleration value for the new location and calibrate the scale according to the procedure above. Refer to the previous section to set the gravity acceleration value.

20

Image 22
Contents SW Series Page Contents Options MaintenanceSpecifications Gravity Acceleration MAPCompliance with FCC rules ComplianceClassification of protection provided by enclosures Features IntroductionCleaning the scale PrecautionsInstalling the scale Operating the scaleUnpacking Description of Each Part Display rear sideSymbol Description Display and symbolsSymbols DisplaySwitches Attaching the display pole to the base Setting UPSW-M / L series with a display pole Installing the scale Bat oo Turning the power on and OFFBasic Operation ON/OFFBasic weighing operation Indicator will not turn onSelecting a weighing unit During the Tare operationWeight display resolution Counting Mode PrintFormula to compare Five-level comparator modeComparator LoloThree-level comparator mode Entering the comparator values Five-level comparator mode Three-level comparator modePrint Sample Tare Calibration Gravity acceleration correctionCalibration using a weight Units PrintRestoring the factory set values Functions Setting the parametersTare ON/OFF Function list Key Installing the battery Using the OP-02 SLA batteryOptions Charging the battery Installing OP-03 OP-03 RS-232C / Relay OutputOP-03 Specifications AsciiSolid sate relay ‰ Data example Command mode Stable positive data Page Installing OP-04 OP-04 RS-422OP-04 Specifications Example of connection RS-422 Stable data Maintenance Error codesSpecifications SpecificationsModel SW-6KS SW-15KS SW-15KM SW-30KM Model SW-60KM SW-60KLExternal dimensions SW-60KL / SW-150KL with a display pole Gravity Acceleration MAP Values of gravity at various locationsWorld map