4.PRECAUTIONS

4.1.Installing the scale

Caution

Ground the scale so that the user will not be subject to an electric shock.

Do not handle the AC power cable with wet hands.

The AC plug is not water-resistant. Use an electrical outlet located at a place where the plug will not get wet.

Do not handle the cables carelessly.

Do not install the scale where flammable or corrosive gas is present.

Consider the following conditions to get the most from your scale.

‰Install the scale where the temperature and relative humidity are stable, with no drafts or vibration.

‰Install the scale on a solid and level surface.

‰Do not install the scale in direct sunlight.

‰Do not install the scale near heaters or air conditioners.

‰Do not install the scale near equipment which produces magnetic fields.

‰Do not install the scale in a place where it is possible to be charged with static electricity, or where the relative humidity is lower than 45% RH. Insulators such as plastic are often charged with static electricity.

‰Use a stable power source.

‰Calibrate the scale before use or after having moved it to another location. In addition, calibrate it periodically to maintain the accuracy.

(Refer to “11. CALIBRATION”).

4.2.Operating the scale

‰Periodically ensure that the scale weighs correctly.

‰Calibrate the scale periodically to maintain the weighing accuracy (Refer to “11. CALIBRATION”).

‰Do not place anything on the pan that is beyond the weighing capacity.

‰Do not apply a shock load to the scale.

‰Touch the switches only with a finger.

‰Confirm zero before each weighing to prevent possible error.

4.3.Cleaning the scale

‰Do not disassemble the scale.

‰Do not use solvents to clean the scale.

‰The scale can be washed with water completely.

5

Image 7
Contents SW Series Page Contents Gravity Acceleration MAP MaintenanceSpecifications OptionsCompliance with FCC rules ComplianceClassification of protection provided by enclosures Introduction FeaturesOperating the scale PrecautionsInstalling the scale Cleaning the scaleUnpacking Display rear side Description of Each PartDisplay Display and symbolsSymbols Symbol DescriptionSwitches Attaching the display pole to the base Setting UPSW-M / L series with a display pole Installing the scale ON/OFF Turning the power on and OFFBasic Operation Bat ooDuring the Tare operation Indicator will not turn onSelecting a weighing unit Basic weighing operationWeight display resolution Print Counting ModeLolo Five-level comparator modeComparator Formula to compareThree-level comparator mode Entering the comparator values Five-level comparator mode Three-level comparator modePrint Sample Tare Gravity acceleration correction CalibrationUnits Print Calibration using a weightRestoring the factory set values Functions Setting the parametersTare ON/OFF Function list Key Installing the battery Using the OP-02 SLA batteryOptions Charging the battery OP-03 RS-232C / Relay Output Installing OP-03Ascii OP-03 SpecificationsSolid sate relay ‰ Data example Command mode Stable positive data Page OP-04 RS-422 Installing OP-04OP-04 Specifications Example of connection RS-422 Stable data Error codes MaintenanceSpecifications SpecificationsModel SW-6KS SW-15KS SW-15KM SW-30KM SW-60KL Model SW-60KMExternal dimensions SW-60KL / SW-150KL with a display pole Values of gravity at various locations Gravity Acceleration MAPWorld map