Lifescan Diabetes Monitoring System manual Important Control Solution Test Information

Page 24

IMPORTANT CONTROL SOLUTION TEST INFORMATION

Use only ONE TOUCHNormal Control Solution.

Check the expiration date of the control solution. Do not use if expired.

Record the discard date on the control solution vial. Discard the vial three months after first opening.

The ONE TOUCHNormal Control Range printed on the test strip vial is for ONE TOUCH Normal Control Solution only. It is not a recom- mended range for your blood glucose level.

Store the control solution at room temperature below 30°C (86°F). Do not refrigerate.

Control solution should not be used as a cleaning solution for the meter.

Warning

For external use only. Do not swallow or inject control solution or put control solution drops into your eyes.

*For detailed information on ONE TOUCHControl Solution, refer to the ONE TOUCHControl Solution insert.

17

Image 24
Contents Owner’s Booklet Dear ONE Touch TM Basic TM Plus Owner Page Page Table of Contents Using the Meter Memory Solving Problems Precautions and Limitations Infection Control Getting Started Page ONE Touch Basic Plus Blood Glucose Meter ButtonPage ONE Touch Test Strips Important Test Strip Information Before Testing Setting Meter CodePress the On/Off Button Match the Code NumbersChecking the System Code the MeterChecking with the Check Strip When insert code 10 Example Press the On/Off Button. Insert Side 1 of the Check StripStep Appears, Remove the Check Strip from the Test Strip HolderExample Checking with Control Solution Page Important Control Solution Test Information When insert code10 Press On/Off Button. Insert Test StripApply Control Solution Control Solution Result Appears in 45 SecondsControl solution results Detailed Test Procedure Penlet Plus Adjustable Blood SamplerGetting a Drop of Blood Page Cock the Penlet Plus Sampler Wash Your HandsLance Your Finger Remove the LancetAlways use caution when removing Check code and then Remove a test strip Apply Blood Sample Accurate Results in 45 Seconds Special Messages Checking the Amount of Blood on the Test Strip Page Using the Meter Memory Entering the Memory Mode Recalling Test Results Setting Meter Options Entering and Using the Set-Up ModePage Clock Time and Date Setting the Time Beep Prompts Language Time Format Date FormatUnit of Measure Decimal SeparatorExiting the Set-Up Mode Solving Problems Test Procedure MessagesCode Check code Insert Code 1 0 Example Strip wait Apply sampleInsert s i d e Alternating with Error Messages Battry Clean test area Remove stripError 1 retest Error 2 retestMessage Problem What to do Error3 retest Not enough blood retestT o k D o a RetestMeter requires a Taking Care Your Meter Cleaning the Meter Doing a Daily CheckClean the Test Strip Holder Remove the Test Strip Holder from the MeterClean the Test Area Do not USEReplace the Test Strip Holder Replacing the Batteries To replace the batteries References Specifications Result Range 0-600 mg/dLWarranty Three-Year WarrantyPage Page IFD/OB/OTBP