Step 2: Apply Control Solution.
Shake the control
Step 3: Control Solution Result Appears in 45 Seconds.
Plussolution vial well before using. If there are any bubbles on
M the top of the vial,
wipe them off with a
clean tissue. Remove the cap and apply a
drop of control solution to the middle of the test spot, covering it completely. Do not dab, wipe, or
Compare the control solution test result with the ONE TOUCH™ Normal Control Range printed on the test strip vial. When you get a control solution test result within the control range, you are ready to test your blood.
M | C |
(Example) |
smear control solution on the test spot. Avoid contaminating the control solution. Do not touch the test spot with the tip of the vial.
Note: The control solution range is shown in two different units of
ONE TOUCH™ Normal Control Range
(Example)
19