Leica Leica VT1200S manual Working with the Instrument

Page 25

 

5.

Working with the Instrument

 

 

 

Key/setting dial

semiautomatic sectioning mode

automatic sectioning mode

 

 

 

 

The rotary knob for moving the

ditto

 

 

specimen receptacle vertically can

 

2

be used for fine-adjusting the

 

blade-to-specimen proximity. Turn-

 

 

 

 

ing the setting dial clockwise

 

 

 

 

moves the specimen up into the

 

 

desired proximity; turning the set-

 

 

ting dial counterclockwise lowers

 

 

the specimen (minus sign).

 

 

The step sizes 1, 10 or 100μm can

 

 

be selected by pressing the STEP

 

 

SIZE key.

 

 

After the setting dial is turned

 

 

clockwise or counter clockwise,

 

 

the specimen receptacle position is

 

 

updated in the μm display.

 

 

In semiautomatic mode, the de-

Not possible.

 

sired section thickness is selected

 

 

using the setting dial.

 

 

The selected section thickness

 

 

appears in the μm display and the

 

 

current specimen receptacle posi-

 

 

tion appears in the μm display.

 

 

After each sectioning process has

 

 

been completed, the μm display is

 

 

set to "0".

 

 

 

 

 

When the DOWN key is pressed,

ditto

 

the specimen receptacle automat-

 

 

ically moves quickly into the low-

 

 

est position. (The LED in the DOWN

 

 

key lights up when the lower end

 

 

position is reached.) The μm dis-

 

 

play is set to "0".

 

Leica VT1200 and VT1200S – Microtome

25

Image 25
Contents Leica VT1200 Leica VT1200S Page Nussloch Germany Contents Important Notes Symbols in the Text and their MeaningsIntended Use/Improper Use Qualification of PersonnelGeneral Safety Notes SafetyProper handling Transport, Unpacking and Setting upTechnical Data for the VT1200 Instrument CharacteristicsElectrical data Technical Data for the VT1200S Instrument CharacteristicsDimensions Ambient conditionsMicroscope General Overview of the VT1200/VT1200SBlades for the blade holder Attachments on dovetail receptacleLED for feed rate VT1200 Control PanelWindow edge is set Moves the specimen Upwards as long as 1 VT1200S Control PanelStandard Scope of Delivery for the VT1200 Installation0481 Standard Scope of Delivery for the VT1200S91200S001 Packing the instrument Packing and Setting up the InstrumentBefore Commissioning the Instrument Description of the Typical Application Working with the InstrumentMoving the Specimen Receptacle Vertically Control Elements on the VT1200 Control PanelBlade Forward back Blade MethodStarts the sectioning process Selecting the Sectioning Parameters Control Element on the Control Panel of the VT1200S LED illumination is switched on Blade feed rate Speed Selected amplitude Ampl Saved section thicknessWhen these keys are pressed Toggling from MAN to An audible warning signal soundsToggling from Auto ModeKey/setting dial Adjusting the amplitude fromAutomatic mode max μm Mm in 0.05 mm increments Not possibleWorking with the Instrument Blade forward and Blade Ward key. Press the 1st cutting Key lights upWindow edge key until the LED Ward key and press 2nd cuttingProcess Μm display is then reset toUsing predefined step sizes 1 Or 100μm by max μm using rotary knobAutomatic section thickness Stroke type CUT only single MAN with rotary knob 1 forStroke Single can be selected With rotary knob 1. If continuous5 mm/s in 0.5 mm/s increments Installing the Ice Bath and Buffer Tray Installing the AccessoriesPreparing a Specimen Installing the Double-walled Buffer TrayInstalling and Removing the Blade Holder Now return the blade holder to the cutting position Inserting the Blade Adjusting the Clearance AngleWorking with the Instrument Operating the VibroCheck Using the VibroCheck with the VT1200Normal operating status is now restored Using the VibroCheck with the VT1200S Operating the VibroCheck Error Messages and Troubleshooting Malfunctions Meanings and TroubleshootingError/INF No Troubleshooting Remark Harm the instrumentReached InF Software updatedPanel and keyboard C2 have differentDown during installa Dirty. CleanVibroCheck VibroCheck was Collides with BladeNot accepted Negative values Values of more than 1000μm Blocked when RUN is activeInstrument is Plug connections are Replacing the Main Fuse Cleaning the Instrument Cleaning and MaintenanceClean using an alcohol-based solution or acetone Specimen plate Buffer trayBlade DisplayBuffer Tray Extension Accessories for Standard SpecimensLid for covering the buffer tray Specimen PlateVibroCheck Magnifier, Complete Blades Microscope, CompleteInstalling the Magnifier Support or Microscope Support 230-240 V, 50/60 Hz Fiber-optic Illumination, Cold Light Source240 V, 50/60 Hz Voltage variant 230V/50/60Hz Leica VT1200/VT1200S Microtome with vibrating blade EC Declaration of ConformityWarranty Warranty and ServiceOperating Manual V1.1 05/2007