3.1 | Technical Data for the VT1200 |
|
General data: |
| |
| Cutting frequency (± 10 %) | 85 Hz (± 10%) |
| Amplitude | from 0 – 3mm, in 0.05mm increments |
| Cutting speed (± 10 %) | 0.01 – 1.5mm/s |
| Return speed (± 10 %) | 2.5mm/s |
| Total vertical specimen stroke | 20 mm (motorized) |
| Cutting range | 45 mm (adjustable) |
| Maximum specimen size: |
|
| With standard blade holder ............................................................................................................. 33 x 50 mm | |
| Specimen orientation, rotating ................................................................................................................... 360° | |
| Specimen plate, swiveling ...................................................................................................................... 0 – 10° | |
| Section thickness adjustment | manual, in 1 μm increments |
Ambient conditions: |
| |
| Operating temperature range: | min. 10 °C – max. 35 °C |
| Relative humidity: | max. 60 % |
| Storage temperature: | 5 – 55°C |
| Storage humidity: | < 60 % |
| Height: | up to 2000 m above sea level |
Electrical data: |
| |
| Rated voltage range (± 10 %): | 100V – 240V |
| Rated frequency (± 10 %): | 50/60Hz |
| Power consumption: | 35 VA |
| Protection class: | I |
| Power fuses: | T 1A L 250 V |
| Pollution degree: | 2 |
| Overvoltage category: | II |
| Electrical overload protection: | yes |
| Internal current limit for the electronics: | yes |
Dimensions: |
| |
| L x W x H: | 600mm x 250mm x 230mm |
| Height with magnifier support | 600mm x 250mm x 320mm |
| Height with microscope: | 600mm x 250mm x 490mm |
| L x W x H control unit (when the bases are folded in): | 165mm x 120mm x 72mm |
| Weight: |
|
| Without magnifier support and control unit | 56 kg |
| VT1200 control unit | 1 kg |
| Magnifier support | 2 kg |
| Microscope support with stereomicroscope | 4 kg |