7. Malfunctions: Meanings and Troubleshooting

Error/INF No.

Error

Troubleshooting

Remark

Error 52

-

Basic calibration of

-

Emitter diode or receiver

 

 

 

light barrier is not pos-

 

is defective.

 

 

 

sible for VibroCheck.

-

VibroCheck is defective.

 

 

 

 

-

Inform service!

 

 

 

 

 

 

 

Error 53

-

VibroCheck blade

-

No blade and/or blade

- VT1200 drives to the

 

 

search not successful.

 

holder swung out or not

lowest Z position us-

 

 

 

 

installed.

ing the stepper motor

 

 

 

-

Emitter diode or receiver

after an unsuccessful

 

 

 

 

dirty. Clean

search. (Allows blade

 

 

 

-

VibroCheck is defective.

installation)

 

 

 

 

 

 

Communication

-

Communication error

-

VT1200S: display lights up,

 

between keyboard

 

between control unit

 

but remains blank.

 

– VT1200/S

 

and VT.

-

VT1200: row of dots

 

 

 

 

 

flashes.

 

 

 

 

-

Inform service!

 

 

 

 

 

 

 

- The knife/blade

- The buffer tray was not

collides with the

pushed in up to the

buffer tray

rear pin or was not

 

clamped down during

 

installation.

Push in the buffer tray as far as it will go and clamp it down with the lever.

- The VibroCheck

- The VibroCheck was

collides with the

not pushed all the way

blade

in or was not clamped

 

down during installa-

 

tion.

-Push in the VibroCheck as far as it will go and clamp it down with the lever.

-Switching from SINGL - In MAN mode, only single cuts are possible. to CONT is not possible.

- The cutting window

-

In MAN mode, only single cuts are possible.

cannot be defined.

-

If multiple cuts are required, switch to AUTO mode.

44

Operating Manual V1.0 – 10/2006

Image 44
Contents Leica VT1200 Leica VT1200S Page Nussloch Germany Contents Symbols in the Text and their Meanings Important NotesIntended Use/Improper Use Qualification of PersonnelSafety General Safety NotesTransport, Unpacking and Setting up Proper handlingElectrical data Instrument CharacteristicsTechnical Data for the VT1200 Instrument Characteristics Technical Data for the VT1200SDimensions Ambient conditionsGeneral Overview of the VT1200/VT1200S MicroscopeAttachments on dovetail receptacle Blades for the blade holderVT1200 Control Panel LED for feed rate1 VT1200S Control Panel Window edge is set Moves the specimen Upwards as long asInstallation Standard Scope of Delivery for the VT120091200S001 Standard Scope of Delivery for the VT1200S0481 Packing and Setting up the Instrument Packing the instrumentBefore Commissioning the Instrument Working with the Instrument Description of the Typical ApplicationControl Elements on the VT1200 Control Panel Moving the Specimen Receptacle VerticallyBlade Method Blade Forward backSelecting the Sectioning Parameters Starts the sectioning processLED illumination is switched on Control Element on the Control Panel of the VT1200SBlade feed rate Speed Selected amplitude Ampl Saved section thicknessToggling from MAN to An audible warning signal sounds When these keys are pressedToggling from Auto ModeAdjusting the amplitude from Key/setting dialAutomatic mode max μm Mm in 0.05 mm increments Not possibleWorking with the Instrument Blade forward and Blade Key lights up Ward key. Press the 1st cuttingWindow edge key until the LED Ward key and press 2nd cuttingΜm display is then reset to ProcessAutomatic section thickness Or 100μm by max μm using rotary knobUsing predefined step sizes 1 MAN with rotary knob 1 for Stroke type CUT only singleStroke Single can be selected With rotary knob 1. If continuous5 mm/s in 0.5 mm/s increments Installing the Accessories Installing the Ice Bath and Buffer TrayInstalling the Double-walled Buffer Tray Preparing a SpecimenInstalling and Removing the Blade Holder Inserting the Blade Adjusting the Clearance Angle Now return the blade holder to the cutting positionWorking with the Instrument Using the VibroCheck with the VT1200 Operating the VibroCheckNormal operating status is now restored Using the VibroCheck with the VT1200S Operating the VibroCheck Malfunctions Meanings and Troubleshooting Error Messages and TroubleshootingError/INF No Troubleshooting Remark Harm the instrumentReached Software updated InFPanel and keyboard C2 have differentDirty. Clean Down during installaVibroCheck VibroCheck was Collides with BladeInstrument is Plug connections are Blocked when RUN is activeNot accepted Negative values Values of more than 1000μm Replacing the Main Fuse Clean using an alcohol-based solution or acetone Cleaning and MaintenanceCleaning the Instrument Buffer tray Specimen plateBlade DisplayExtension Accessories for Standard Specimens Buffer TraySpecimen Plate Lid for covering the buffer trayVibroCheck Blades Microscope, Complete Magnifier, CompleteInstalling the Magnifier Support or Microscope Support 240 V, 50/60 Hz Fiber-optic Illumination, Cold Light Source230-240 V, 50/60 Hz Voltage variant 230V/50/60Hz EC Declaration of Conformity Leica VT1200/VT1200S Microtome with vibrating bladeWarranty and Service WarrantyOperating Manual V1.1 05/2007