6. Operating the VibroCheck

6.1Using the VibroCheck with the VT1200

The following instructions must be adhered to exactly. Noncompliance can cause serious damage to the instrument.

We recommend using the VibroCheck after each blade change to check the optimum position of the blade and minimize the vertical vibration.

Foot switch for VT

(2) – VibroCheck

Control panel for VT

Fig. 23

3

Fig. 24

15

Prior to assembly, the dovetail guide (3) is brought down to the lowest position using the DOWN key!

1.The user installs the VibroCheck: the VC is pushed along the dovetail guide (3) past the mark on the base plate of the instrument (rear stop) and clamped firmly using the lever (15). Insert the blade and clamp tightly. Return the blade to the cutting position (see Fig. 20).

2. Plug the connecting plug of the VibroCheck (VC) into the socket (2) on the left side panel. Short red flash of LED on VC --> control panel acknowledges VC. LED then remains red. LED in DOWN key flashes green.

3.The user presses the DOWN key. The VC moves to the bottom position, after that the blade moves to the rearmost position – the LED in the RUN/STOP key flashes. Short red flash of LED on VC --> it then remains red.

4. The user presses the RUN/STOP key: the blade first moves towards the front (into the posi- tion exactly above the VibroCheck); the VC then moves into a position in which the blade partly covers the light barrier. (If the VC does not receive any signal via the light barrier, the operation is canceled and DOWN is enabled.) LED on VC flashes green – RUN/STOP flashes yellow.

Fig. 25

Leica VT1200 and VT1200S – Microtome

37

Image 37
Contents Leica VT1200 Leica VT1200S Page Nussloch Germany Contents Important Notes Symbols in the Text and their MeaningsIntended Use/Improper Use Qualification of PersonnelGeneral Safety Notes SafetyProper handling Transport, Unpacking and Setting upTechnical Data for the VT1200 Instrument CharacteristicsElectrical data Technical Data for the VT1200S Instrument CharacteristicsDimensions Ambient conditionsMicroscope General Overview of the VT1200/VT1200SBlades for the blade holder Attachments on dovetail receptacleLED for feed rate VT1200 Control PanelWindow edge is set Moves the specimen Upwards as long as 1 VT1200S Control PanelStandard Scope of Delivery for the VT1200 Installation0481 Standard Scope of Delivery for the VT1200S91200S001 Packing the instrument Packing and Setting up the InstrumentBefore Commissioning the Instrument Description of the Typical Application Working with the InstrumentMoving the Specimen Receptacle Vertically Control Elements on the VT1200 Control PanelBlade Forward back Blade MethodStarts the sectioning process Selecting the Sectioning ParametersControl Element on the Control Panel of the VT1200S LED illumination is switched onBlade feed rate Speed Selected amplitude Ampl Saved section thicknessWhen these keys are pressed Toggling from MAN to An audible warning signal soundsToggling from Auto ModeKey/setting dial Adjusting the amplitude fromAutomatic mode max μm Mm in 0.05 mm increments Not possibleWorking with the Instrument Blade forward and Blade Ward key. Press the 1st cutting Key lights upWindow edge key until the LED Ward key and press 2nd cuttingProcess Μm display is then reset toUsing predefined step sizes 1 Or 100μm by max μm using rotary knobAutomatic section thickness Stroke type CUT only single MAN with rotary knob 1 forStroke Single can be selected With rotary knob 1. If continuous5 mm/s in 0.5 mm/s increments Installing the Ice Bath and Buffer Tray Installing the AccessoriesPreparing a Specimen Installing the Double-walled Buffer TrayInstalling and Removing the Blade Holder Now return the blade holder to the cutting position Inserting the Blade Adjusting the Clearance AngleWorking with the Instrument Operating the VibroCheck Using the VibroCheck with the VT1200Normal operating status is now restored Using the VibroCheck with the VT1200S Operating the VibroCheck Error Messages and Troubleshooting Malfunctions Meanings and TroubleshootingError/INF No Troubleshooting Remark Harm the instrumentReached InF Software updatedPanel and keyboard C2 have differentDown during installa Dirty. CleanVibroCheck VibroCheck was Collides with BladeNot accepted Negative values Values of more than 1000μm Blocked when RUN is activeInstrument is Plug connections are Replacing the Main Fuse Cleaning the Instrument Cleaning and MaintenanceClean using an alcohol-based solution or acetone Specimen plate Buffer trayBlade DisplayBuffer Tray Extension Accessories for Standard SpecimensLid for covering the buffer tray Specimen PlateVibroCheck Magnifier, Complete Blades Microscope, CompleteInstalling the Magnifier Support or Microscope Support 230-240 V, 50/60 Hz Fiber-optic Illumination, Cold Light Source240 V, 50/60 Hz Voltage variant 230V/50/60Hz Leica VT1200/VT1200S Microtome with vibrating blade EC Declaration of ConformityWarranty Warranty and ServiceOperating Manual V1.1 05/2007