7.Malfunctions: Meanings and Troubleshooting
7.1Error Messages and Troubleshooting
Error/INF No. | Error | Troubleshooting | Remark | ||
|
|
|
|
|
|
Error 01 | - | Wrong control panel | - | Use the correct control | - Using an incorrect |
|
| (VT1200 or VT1200S) |
| panel for the instrument. | control panel does not |
|
|
|
|
| harm the instrument, |
|
|
|
|
| but no functions what- |
|
|
|
|
| soever are available. |
|
|
|
|
| |
Error 21 | - | Head does not swing | 1. Check to see if an ampli- |
| |
|
| into place. |
| tude value has been se- |
|
|
|
|
| lected. |
|
|
|
| 2. | Try to make the head |
|
|
|
|
| swing into place by gently |
|
|
|
|
| striking it with your hand. |
|
|
|
| 3. | If head does not swing |
|
|
|
|
| into place, inform service. |
|
|
|
|
|
|
|
Error 22 | - | Timeout while initializ- | - | Inform service! |
|
|
| ing the |
|
|
|
|
|
|
|
| |
Error 23 | - DC motor of | - | Inform service! |
| |
|
| not rotate (during ini- |
|
|
|
|
| tialization or in normal |
|
|
|
|
| operation) |
|
|
|
|
|
|
|
| |
Error 24 | - | - | Inform service! |
| |
|
| not reached. |
|
|
|
|
|
|
|
|
|
Error 25 | - | - | Inform service! |
| |
|
| cannot be moved away |
|
|
|
|
| from. |
|
|
|
|
|
|
|
| |
Error 26 | - | - | Inform service! |
| |
|
| not reached. |
|
|
|
After all error messages, the instrument MUST be switched off, then on again using the main switch.
Leica VT1200/VT1200S – Microtome | 41 |