Leica Leica VT1200 InF, Panel and keyboard, C2 have different, Software versions, Or blade holder

Page 43

7.Malfunctions: Meanings and Troubleshooting

Error/INF No.

Error

Troubleshooting

Remark

InF 41

- Control board (C1/back

 

panel) and keyboard

 

(C2) have different

 

software versions.

-It is possible that due to different software ver- sions, some or all func- tions of the instrument may be partly or entirely unavailable.

Inform service and have the software updated to the latest version.

InF 42

- Control board (C1/back

 

panel) and VibroCheck

 

(C3) have different

 

software versions.

-It is possible that due to different software ver- sions, some or all func- tions of the instrument may be partly or entirely unavailable.

"InF" messages can be removed by pressing the Clear key.

The message appears once each time the instrument is switched ON. "InF" messages do not cause the instrument to be locked.

After all error messages, the instrument MUST be switched off, then on again using the main switch. – Exceptions are InF 41 and 42.

Error 51

- Horizontal calibration

- Possible operating error,

 

of VibroCheck is not

see Operating Manual.

 

possible.

- Great deal of damage or

 

 

dirt on the blade or blade

 

 

holder. Use a new blade

 

 

or blade holder.

 

 

- Controllers have different

 

 

software versions. Have

 

 

the software updated.

 

 

- VibroCheck is defective.

 

 

Have it checked by ser-

 

 

vice.

- A calibration is car- ried out even if the blade is jagged.

A great deal of obvi- ous damage or dirt must be present before a blade cali- bration is canceled automatically.

Leica VT1200/VT1200S – Microtome

43

Image 43
Contents Leica VT1200 Leica VT1200S Page Nussloch Germany Contents Qualification of Personnel Symbols in the Text and their MeaningsImportant Notes Intended Use/Improper UseGeneral Safety Notes SafetyProper handling Transport, Unpacking and Setting upTechnical Data for the VT1200 Instrument CharacteristicsElectrical data Ambient conditions Instrument CharacteristicsTechnical Data for the VT1200S DimensionsMicroscope General Overview of the VT1200/VT1200SBlades for the blade holder Attachments on dovetail receptacleLED for feed rate VT1200 Control PanelWindow edge is set Moves the specimen Upwards as long as 1 VT1200S Control PanelStandard Scope of Delivery for the VT1200 Installation0481 Standard Scope of Delivery for the VT1200S91200S001 Packing the instrument Packing and Setting up the InstrumentBefore Commissioning the Instrument Description of the Typical Application Working with the InstrumentMoving the Specimen Receptacle Vertically Control Elements on the VT1200 Control PanelBlade Forward back Blade MethodStarts the sectioning process Selecting the Sectioning ParametersSelected amplitude Ampl Saved section thickness LED illumination is switched onControl Element on the Control Panel of the VT1200S Blade feed rate SpeedMode Toggling from MAN to An audible warning signal soundsWhen these keys are pressed Toggling from AutoMm in 0.05 mm increments Not possible Adjusting the amplitude fromKey/setting dial Automatic mode max μmWorking with the Instrument Blade forward and Blade Ward key and press 2nd cutting Key lights upWard key. Press the 1st cutting Window edge key until the LEDProcess Μm display is then reset toUsing predefined step sizes 1 Or 100μm by max μm using rotary knobAutomatic section thickness With rotary knob 1. If continuous MAN with rotary knob 1 forStroke type CUT only single Stroke Single can be selected5 mm/s in 0.5 mm/s increments Installing the Ice Bath and Buffer Tray Installing the AccessoriesPreparing a Specimen Installing the Double-walled Buffer TrayInstalling and Removing the Blade Holder Now return the blade holder to the cutting position Inserting the Blade Adjusting the Clearance AngleWorking with the Instrument Operating the VibroCheck Using the VibroCheck with the VT1200Normal operating status is now restored Using the VibroCheck with the VT1200S Operating the VibroCheck Harm the instrument Malfunctions Meanings and TroubleshootingError Messages and Troubleshooting Error/INF No Troubleshooting RemarkReached C2 have different Software updatedInF Panel and keyboardBlade Dirty. CleanDown during installa VibroCheck VibroCheck was Collides withNot accepted Negative values Values of more than 1000μm Blocked when RUN is activeInstrument is Plug connections are Replacing the Main Fuse Cleaning the Instrument Cleaning and MaintenanceClean using an alcohol-based solution or acetone Display Buffer traySpecimen plate BladeBuffer Tray Extension Accessories for Standard SpecimensLid for covering the buffer tray Specimen PlateVibroCheck Magnifier, Complete Blades Microscope, CompleteInstalling the Magnifier Support or Microscope Support 230-240 V, 50/60 Hz Fiber-optic Illumination, Cold Light Source240 V, 50/60 Hz Voltage variant 230V/50/60Hz Leica VT1200/VT1200S Microtome with vibrating blade EC Declaration of ConformityWarranty Warranty and ServiceOperating Manual V1.1 05/2007