Leica Leica VT1200S manual Normal operating status is now restored

Page 38

6.Operating the VibroCheck

1

2

Fig. 26

17

18a

18

18

16

Fig. 27

If the display shows μm (2) "0" (opti- mal) and the value in μm (1) is unac- ceptably high, the blade must be replaced.

When LED on VC is green and LED in RUN/ STOP is yellow, the blade starts to vibrate with the amplitude set. Speed = 0, amplitude can be changed at any time.

On the (5-digit) display, the deviation in height amplitude is shown in μm (e.g. 0.9μm). This val- ue can be either positive or negative. The dis- play can be switched using the key μm/μm. Then a number appears (e.g. 0.4). This means clockwise rotation by 0.4 turns ("+" see 18a) – (Neg. operational sign means counterclockwise rotation – see 18a). If the value is "0", no improve- ment of the height amplitude is possible.

5. Press the STOP key. Using a size 3 Allen key, unscrew the clamping screw (16) only slightly, pull off the cap (17) by pulling it upwards (keep it in a safe place) and turn the adjusting screw (18) by the corresponding value (here, 0.4 turns) clockwise (in the "+" direction, 18a) using a size

3 Allen key. Tighten the clamping screw (16) clockwise.

6.Press the RUN key, check the value, repeat steps 5 – 7 if necessary.

7.If the measured value is accepted, press DOWN (flashes green). VC moves to lowest position – blade is moved to the rear. LED on VC illuminates red again (LED on RUN/STOP is off).

8.The instrument now expects the VibroCheck to be removed. To do so, disconnect the USB plug connection of the VC to the basic instru- ment and pull the VC off of the dovetail guide. LED in DOWN key continues to flash – wait until LED in DOWN button goes out.

The normal operating status is now restored.

38

Operating Manual V1.1 – 05/2007

Image 38
Contents Leica VT1200 Leica VT1200S Page Nussloch Germany Contents Intended Use/Improper Use Symbols in the Text and their MeaningsImportant Notes Qualification of PersonnelSafety General Safety NotesTransport, Unpacking and Setting up Proper handlingElectrical data Instrument CharacteristicsTechnical Data for the VT1200 Dimensions Instrument CharacteristicsTechnical Data for the VT1200S Ambient conditionsGeneral Overview of the VT1200/VT1200S MicroscopeAttachments on dovetail receptacle Blades for the blade holderVT1200 Control Panel LED for feed rate1 VT1200S Control Panel Window edge is set Moves the specimen Upwards as long asInstallation Standard Scope of Delivery for the VT120091200S001 Standard Scope of Delivery for the VT1200S0481 Packing and Setting up the Instrument Packing the instrumentBefore Commissioning the Instrument Working with the Instrument Description of the Typical ApplicationControl Elements on the VT1200 Control Panel Moving the Specimen Receptacle VerticallyBlade Method Blade Forward backSelecting the Sectioning Parameters Starts the sectioning processBlade feed rate Speed LED illumination is switched onControl Element on the Control Panel of the VT1200S Selected amplitude Ampl Saved section thicknessToggling from Auto Toggling from MAN to An audible warning signal soundsWhen these keys are pressed ModeAutomatic mode max μm Adjusting the amplitude fromKey/setting dial Mm in 0.05 mm increments Not possibleWorking with the Instrument Blade forward and Blade Window edge key until the LED Key lights upWard key. Press the 1st cutting Ward key and press 2nd cuttingΜm display is then reset to ProcessAutomatic section thickness Or 100μm by max μm using rotary knobUsing predefined step sizes 1 Stroke Single can be selected MAN with rotary knob 1 forStroke type CUT only single With rotary knob 1. If continuous5 mm/s in 0.5 mm/s increments Installing the Accessories Installing the Ice Bath and Buffer TrayInstalling the Double-walled Buffer Tray Preparing a SpecimenInstalling and Removing the Blade Holder Inserting the Blade Adjusting the Clearance Angle Now return the blade holder to the cutting positionWorking with the Instrument Using the VibroCheck with the VT1200 Operating the VibroCheckNormal operating status is now restored Using the VibroCheck with the VT1200S Operating the VibroCheck Error/INF No Troubleshooting Remark Malfunctions Meanings and TroubleshootingError Messages and Troubleshooting Harm the instrumentReached Panel and keyboard Software updatedInF C2 have differentVibroCheck VibroCheck was Collides with Dirty. CleanDown during installa BladeInstrument is Plug connections are Blocked when RUN is activeNot accepted Negative values Values of more than 1000μm Replacing the Main Fuse Clean using an alcohol-based solution or acetone Cleaning and MaintenanceCleaning the Instrument Blade Buffer traySpecimen plate DisplayExtension Accessories for Standard Specimens Buffer TraySpecimen Plate Lid for covering the buffer trayVibroCheck Blades Microscope, Complete Magnifier, CompleteInstalling the Magnifier Support or Microscope Support 240 V, 50/60 Hz Fiber-optic Illumination, Cold Light Source230-240 V, 50/60 Hz Voltage variant 230V/50/60Hz EC Declaration of Conformity Leica VT1200/VT1200S Microtome with vibrating bladeWarranty and Service WarrantyOperating Manual V1.1 05/2007