Leica Leica VT1200S manual 5 mm/s in 0.5 mm/s increments

Page 31

 

5.

Working with the Instrument

 

 

 

Key/setting dial

semiautomatic sectioning mode

automatic sectioning mode

 

 

 

 

Rotary knob 2

ditto

1

The feed rate (FOR/REV) for the

 

 

"Blade forward" and the "Blade

 

 

back" keys can be set between 1

 

2

and 5 mm/s in 0.5 mm/s increments

 

 

 

 

with rotary knob 1.

 

 

 

 

1

Rotary knob 2

ditto

Switches the vibration movement

 

 

 

 

(FOR/VIB) of the "Blade forward"

 

 

key either on or off using rotary

 

2

knob 1.

 

 

 

 

 

 

ditto

1

Rotary knob 2

 

 

Key acknowledgment (BEEP) on or

 

 

off with rotary knob 1.

 

2

 

 

 

 

 

 

Save the parameter and leave by

ditto

 

pressing the menu MENU key.

 

Leaving the menu and saving the parameter is possible at any point in the menu.

If you want to call up the parameters of a certain user (for example, user 3), press the Menu key and select user 3. Then confirm by pressing the Menu key 2x. The parameters stored under user3 are now enabled.

Leica VT1200 and VT1200S – Microtome

31

Image 31
Contents Leica VT1200 Leica VT1200S Page Nussloch Germany Contents Qualification of Personnel Symbols in the Text and their MeaningsImportant Notes Intended Use/Improper UseGeneral Safety Notes SafetyProper handling Transport, Unpacking and Setting upTechnical Data for the VT1200 Instrument CharacteristicsElectrical data Ambient conditions Instrument CharacteristicsTechnical Data for the VT1200S DimensionsMicroscope General Overview of the VT1200/VT1200SBlades for the blade holder Attachments on dovetail receptacleLED for feed rate VT1200 Control PanelWindow edge is set Moves the specimen Upwards as long as 1 VT1200S Control PanelStandard Scope of Delivery for the VT1200 Installation0481 Standard Scope of Delivery for the VT1200S91200S001 Packing the instrument Packing and Setting up the InstrumentBefore Commissioning the Instrument Description of the Typical Application Working with the InstrumentMoving the Specimen Receptacle Vertically Control Elements on the VT1200 Control PanelBlade Forward back Blade MethodStarts the sectioning process Selecting the Sectioning ParametersSelected amplitude Ampl Saved section thickness LED illumination is switched onControl Element on the Control Panel of the VT1200S Blade feed rate SpeedMode Toggling from MAN to An audible warning signal soundsWhen these keys are pressed Toggling from AutoMm in 0.05 mm increments Not possible Adjusting the amplitude fromKey/setting dial Automatic mode max μmWorking with the Instrument Blade forward and Blade Ward key and press 2nd cutting Key lights upWard key. Press the 1st cutting Window edge key until the LED Process Μm display is then reset toUsing predefined step sizes 1 Or 100μm by max μm using rotary knobAutomatic section thickness With rotary knob 1. If continuous MAN with rotary knob 1 forStroke type CUT only single Stroke Single can be selected5 mm/s in 0.5 mm/s increments Installing the Ice Bath and Buffer Tray Installing the AccessoriesPreparing a Specimen Installing the Double-walled Buffer TrayInstalling and Removing the Blade Holder Now return the blade holder to the cutting position Inserting the Blade Adjusting the Clearance AngleWorking with the Instrument Operating the VibroCheck Using the VibroCheck with the VT1200Normal operating status is now restored Using the VibroCheck with the VT1200S Operating the VibroCheck Harm the instrument Malfunctions Meanings and TroubleshootingError Messages and Troubleshooting Error/INF No Troubleshooting RemarkReached C2 have different Software updatedInF Panel and keyboardBlade Dirty. CleanDown during installa VibroCheck VibroCheck was Collides withNot accepted Negative values Values of more than 1000μm Blocked when RUN is activeInstrument is Plug connections are Replacing the Main Fuse Cleaning the Instrument Cleaning and MaintenanceClean using an alcohol-based solution or acetone Display Buffer traySpecimen plate BladeBuffer Tray Extension Accessories for Standard SpecimensLid for covering the buffer tray Specimen PlateVibroCheck Magnifier, Complete Blades Microscope, CompleteInstalling the Magnifier Support or Microscope Support 230-240 V, 50/60 Hz Fiber-optic Illumination, Cold Light Source240 V, 50/60 Hz Voltage variant 230V/50/60Hz Leica VT1200/VT1200S Microtome with vibrating blade EC Declaration of ConformityWarranty Warranty and ServiceOperating Manual V1.1 05/2007