Complete Condenser Centering
If you have purchased a standard Leica
DM750 configuration, the condenser has already been
tems.
1.Open the condenser aperture by rotating the knurled ring on the condenser to the right.
Make sure that the condenser is in the highest position.
2.Write an "X" on a piece of paper in the size of a business card and place it onto the light output of the microscope stand in such a way that the "X" is centered over the illumination.
Not centered
3.Look at the X through the eyepieces and center it in the field of view by tightening
the setscrews with the tool included in the | Centered | |
delivery package (or, for a Leica DM750 | ||
| ||
with Koehler illumination, tighten the |
| |
knurled screws). |
|
Leica DM750 | Manual | 32 |
|
|
|