Emerson 6081-P instruction manual Section Installation, Considerations, Unpacking and Inspection

Page 15

MODEL 6081-P pH/ORP

SECTION 2.0

 

INSTALLATION

SECTION 2.0

INSTALLATION

2.1Considerations

2.2Unpacking and Inspection

2.3Pre-Installation Set Up

2.4Mechanical Installation

2.5Ground the Transmitter

2.6Power Module Installation

2.1 CONSIDERATIONS

The transmitter can be commissioned before or after installation. It may be useful to commission it on the bench, before installation, to ensure proper operation and to become familiar with its functionality. When applicable, make sure the instru- ments are installed in accordance with intrinsically safe or non-incendive field wiring practices. The device will be powered whenever the power module is installed. To avoid depleting the power module, make sure it is removed when the device is not in use.

Power module

The Model 6081 is battery powered. The power module with the wireless unit contains 2 “C” size primary lithium/thionyl chloride batteries. Each power module contains approximately .5 grams of lithium. Under normal conditions, the power module materials are self-contained and are not reactive as long as the power module integrity is maintained. Care should be taken to prevent thermal, electrical or mechanical damage. Contacts should be protected to prevent premature discharge. Use caution when handling the power module. The power module may be damaged if dropped from heights in excess of 20 feet.

Sensor

Make sensor connections through the cable entry in the enclosure. Be sure to provide adequate clearance for cover removal.

Environmental

Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications.

2.2 UNPACKING AND INSPECTION

Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, remove the transmitter. Be sure all items shown on the packing list are present. If items are missing, immediately notify Rosemount Analytical.

Save the shipping container and packaging. They can be reused if it is later necessary to return the transmitter to the factory.

5

Image 15
Contents Model 6081-P Essential Instructions Shipping considerations for wireless products Power ModulesAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D List of Figures Number TitleIii This page left blank intentionally Section Description and Specifications Features and ApplicationsSpecifications General PH Range 0 to Specifications WirelessFunctional Specifications Measurement update rate 1/sec. to 1/10 minHart Communications Telecommunication ComplianceFCC and IC Asset Management Solutions Section Installation ConsiderationsUnpacking and Inspection Preamplifier Location PRE-INSTALLATION SetupTemperature Element Reference Electrode ImpedanceSection Mounting on a Flat Surface Pipe Mounting Installing the Power Module Power Module InstallationSection This page left blank intentionally Section Sensor Wiring Sensor WiringGeneral Information This page left blank intentionally This page left blank intentionally Section Intrinsically Safe InstallationThis page left blank intentionally Device Network Configuration Section CommissioningNetwork Communications Verify Operation Screen B Section Display and OperationDisplay Screen a Screen CInformation Screen Messages KeypadHow the Security Code Works SecurityThis page left blank intentionally Section Operation with Model Off-line and On-line Operation Making Hart related settings from the keypadChanging START-UP Settings Section Programming the TransmitterGeneral Choosing and Configuring the Analytical Measurement PurposeDefinitions Choose Soln Temp Corr or Sensor Isoptntl Choose MeasurementPurpose Setting a Security Code Procedure Setting a security codeChoose Calib or Config Choose ResetTransmitter Making Hart Related SettingsResetting Factory Calibration and Factory Default Settings Procedure Installing default settingsProcedure Changing screen contrast Selecting a Default Screen and Screen ContrastProcedure Choosing a display screen Choose Default DisplayProcedure Programming the display timeout Choosing a Display TimeoutSection Calibration Temperature Calibrating TemperatureIntroduction Procedure Calibration pH and ORP SectionCalibration Slope and Offset Choose BufferCal Procedure Auto Buffer CalibrationProcedure Manual TWO-POINT Buffer Calibration Choose Standardize Procedure StandardizationProcedure Entering a Known Slope Value ORP Calibration Preparation of ORP standard solutionsThis page left blank intentionally Section Maintenance Transmitter MaintenanceOverview Cleaning Procedures Frequency of CleaningRejuvenating Reference Electrodes Checking the Reference ElectrodeCalibration Frequency ORP Sensor MaintenanceCalibration GeneralPower Module Replacement Handling Considerations Environmental ConsiderationsShipping Considerations General Section Return of MaterialThis page left blank intentionally Warranty Specifications subject to change without notice