Emerson 6081-P instruction manual Procedure Standardization, Choose Standardize

Page 49

MODEL 6081-P pH/ORP

SECTION 10.0

 

CALIBRATION – PH

10.4 PROCEDURE — STANDARDIZATION

1.The pH measured by the transmitter can be changed to match the reading from a second or referee instru- ment. The process of making the two readings agree is called standardization.

2.During standardization, the difference between the two values is converted to the equivalent voltage. The volt- age, called the reference offset, is added to all subsequent measured cell voltages before they are converted to pH. If after standardization the sensor is placed in a buffer solution, the measured pH will differ from the buffer pH by an amount equivalent to the standardization offset.

3.Install the pH sensor in the process liquid.

4.Once readings are stable, measure the pH of the liquid using a referee instrument.

5.Because the pH of the process liquid may change if the temperature changes, measure the pH of the grab sample immediately after taking it.

6.For poorly buffered samples, it is best to determine the pH of a continuously flowing sample from a point as close as possible to the sensor.

7.Press MENU. The main menu appears. Choose Calibrate.

8.Choose pH.

9.Choose Standardize.

10.The top line shows the present reading. Use the arrow keys to change the pH reading in the second line to match the pH reading from the referee instrument.

11.The screen at left appears if the entered pH was greater than 14.00 or if the mV offset calculated by the trans- mitter during standardization exceeds the reference offset limit programmed into the transmitter. The display then returns to step 10. Repeat the standardization. To change the reference offset from the default value (60 mV), see section 8.4.

12.If the entry was accepted the display returns to step 9.

13.To return to the main display, press MENU then EXIT.

39

Image 49
Contents Model 6081-P Essential Instructions Shipping considerations for wireless products Power ModulesAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D Number Title List of FiguresIii This page left blank intentionally Features and Applications Section Description and SpecificationsSpecifications General Functional Specifications Specifications WirelessMeasurement update rate 1/sec. to 1/10 min PH Range 0 toTelecommunication Compliance Hart CommunicationsFCC and IC Asset Management Solutions Considerations Section InstallationUnpacking and Inspection Temperature Element PRE-INSTALLATION SetupReference Electrode Impedance Preamplifier LocationSection Mounting on a Flat Surface Pipe Mounting Installing the Power Module Power Module InstallationSection This page left blank intentionally Sensor Wiring Section Sensor WiringGeneral Information This page left blank intentionally This page left blank intentionally Section Intrinsically Safe InstallationThis page left blank intentionally Section Commissioning Device Network ConfigurationNetwork Communications Verify Operation Display Section Display and OperationScreen a Screen C Screen BInformation Screen Messages KeypadHow the Security Code Works SecurityThis page left blank intentionally Section Operation with Model Off-line and On-line Operation Making Hart related settings from the keypadSection Programming the Transmitter Changing START-UP SettingsGeneral Purpose Choosing and Configuring the Analytical MeasurementDefinitions Choose Soln Temp Corr or Sensor Isoptntl Choose MeasurementPurpose Procedure Setting a security code Setting a Security CodeChoose Calib or Config Resetting Factory Calibration and Factory Default Settings Making Hart Related SettingsProcedure Installing default settings Choose ResetTransmitterProcedure Choosing a display screen Selecting a Default Screen and Screen ContrastChoose Default Display Procedure Changing screen contrastProcedure Programming the display timeout Choosing a Display TimeoutCalibrating Temperature Section Calibration TemperatureIntroduction Procedure Calibration pH and ORP SectionCalibration Slope and Offset Choose BufferCal Procedure Auto Buffer CalibrationProcedure Manual TWO-POINT Buffer Calibration Choose Standardize Procedure StandardizationProcedure Entering a Known Slope Value ORP Calibration Preparation of ORP standard solutionsThis page left blank intentionally Transmitter Maintenance Section MaintenanceOverview Cleaning Procedures Frequency of CleaningRejuvenating Reference Electrodes Checking the Reference ElectrodeCalibration ORP Sensor MaintenanceGeneral Calibration FrequencyPower Module Replacement Environmental Considerations Handling ConsiderationsShipping Considerations General Section Return of MaterialThis page left blank intentionally Warranty Specifications subject to change without notice
Related manuals
Manual 16 pages 28.1 Kb