Emerson 6081-P Section Calibration Temperature, Introduction, Calibrating Temperature

Page 43

MODEL 6081-P pH/ORP

SECTION 9.0

 

CALIBRATION – TEMPERTURE

SECTION 9.0

CALIBRATION — TEMPERATURE

9.1Introduction

9.2Calibrating Temperature

9.1 INTRODUCTION

The Calibrate Menu allows the user to calibrate the pH, ORP (or redox), and temperature response of the sensor.

9.2 CALIBRATING TEMPERATURE

9.2.1 Purpose

Temperature affects the measurement of pH in three ways.

1.The analyzer uses a temperature dependent factor to convert measured cell voltage to pH. Normally, a slight inaccuracy in the temperature reading is unimportant unless the pH reading is significantly different from 7.00. Even then, the error is small. For example, at pH 12 and 25°C, a 1°C error produces a pH error less than ±0.02.

2.During auto calibration, the 6081 recognizes the buffer being used and calculates the actual pH of the buffer at the measured temperature. Because the pH of most buffers changes only slightly with temperature, rea- sonable errors in temperature do not produce large errors in the buffer pH. For example, a 1°C error causes at most an error of ±0.03 in the calculated buffer pH.

3.The 6081 can be programmed to calculate and display pH at a reference temperature (25°C). The maximum change in solution pH with temperature is about ±0.04 pH/°C, so a 1°C temperature error does introduce a small error. However, the major source of error in solution temperature compensation is using an incorrect tem- perature coefficient.

Temperature affects the measurement of ORP in a complicated fashion that is best determined empirically.

Without calibration the accuracy of the temperature measurement is about ±0.4°C. Calibrate the sensor/analyzer combination if

1.±0.4°C accuracy is not acceptable

2.the temperature measurement is suspected of being in error. Calibrate temperature by making the analyzer reading match the temperature measured with a standard thermometer.

33

Image 43
Contents Model 6081-P Essential Instructions Shipping considerations for wireless products Power ModulesAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D Number Title List of FiguresIii This page left blank intentionally Features and Applications Section Description and SpecificationsSpecifications General PH Range 0 to Specifications WirelessFunctional Specifications Measurement update rate 1/sec. to 1/10 minTelecommunication Compliance Hart CommunicationsFCC and IC Asset Management Solutions Considerations Section InstallationUnpacking and Inspection Preamplifier Location PRE-INSTALLATION SetupTemperature Element Reference Electrode ImpedanceSection Mounting on a Flat Surface Pipe Mounting Installing the Power Module Power Module InstallationSection This page left blank intentionally Sensor Wiring Section Sensor WiringGeneral Information This page left blank intentionally This page left blank intentionally Section Intrinsically Safe InstallationThis page left blank intentionally Section Commissioning Device Network ConfigurationNetwork Communications Verify Operation Screen B Section Display and OperationDisplay Screen a Screen CInformation Screen Messages KeypadHow the Security Code Works SecurityThis page left blank intentionally Section Operation with Model Off-line and On-line Operation Making Hart related settings from the keypadSection Programming the Transmitter Changing START-UP SettingsGeneral Purpose Choosing and Configuring the Analytical MeasurementDefinitions Choose Soln Temp Corr or Sensor Isoptntl Choose MeasurementPurpose Procedure Setting a security code Setting a Security CodeChoose Calib or Config Choose ResetTransmitter Making Hart Related SettingsResetting Factory Calibration and Factory Default Settings Procedure Installing default settingsProcedure Changing screen contrast Selecting a Default Screen and Screen ContrastProcedure Choosing a display screen Choose Default DisplayProcedure Programming the display timeout Choosing a Display TimeoutCalibrating Temperature Section Calibration TemperatureIntroduction Procedure Calibration pH and ORP SectionCalibration Slope and Offset Choose BufferCal Procedure Auto Buffer CalibrationProcedure Manual TWO-POINT Buffer Calibration Choose Standardize Procedure StandardizationProcedure Entering a Known Slope Value ORP Calibration Preparation of ORP standard solutionsThis page left blank intentionally Transmitter Maintenance Section MaintenanceOverview Cleaning Procedures Frequency of CleaningRejuvenating Reference Electrodes Checking the Reference ElectrodeCalibration Frequency ORP Sensor MaintenanceCalibration GeneralPower Module Replacement Environmental Considerations Handling ConsiderationsShipping Considerations General Section Return of MaterialThis page left blank intentionally Warranty Specifications subject to change without notice