Emerson 6081-P instruction manual Procedure Entering a Known Slope Value

Page 50

MODEL 6081-P pH/ORP

SECTION 10.0

 

CALIBRATION – PH

10.5 PROCEDURE — ENTERING A KNOWN SLOPE VALUE.

1.If the electrode slope is known from other measurements, it can be entered directly into the transmitter. The slope must be entered as the slope at 25°C. To calculate the slope at 25°C from the slope at temperature t°C, use the equation:

slope at 25°C = (slope at t°C)

298

 

t°C + 273

Changing the slope overrides the slope determined from the previous buffer calibration.

2.Press MENU. The main menu appears. Choose Calibrate.

3.Choose pH.

4.Choose slope.

5.The screen at left appears briefly.

6.Change the slope to the desired value. Press ENTER.

7.The slope must be between 45 and 60 mV/pH. If the value entered is outside this range, the screen at left appears.

8.If the entry was accepted, the screen at left appears.

9.To return to the main display, press MENU then EXIT.

40

Image 50
Contents Model 6081-P Shipping considerations for wireless products Power Modules Essential InstructionsAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D Iii List of FiguresNumber Title This page left blank intentionally Specifications General Section Description and SpecificationsFeatures and Applications Measurement update rate 1/sec. to 1/10 min Specifications WirelessFunctional Specifications PH Range 0 toFCC and IC Hart CommunicationsTelecommunication Compliance Asset Management Solutions Unpacking and Inspection Section InstallationConsiderations Reference Electrode Impedance PRE-INSTALLATION SetupTemperature Element Preamplifier LocationSection Mounting on a Flat Surface Pipe Mounting Power Module Installation Installing the Power ModuleSection This page left blank intentionally General Information Section Sensor WiringSensor Wiring This page left blank intentionally Section Intrinsically Safe Installation This page left blank intentionallyThis page left blank intentionally Network Communications Device Network ConfigurationSection Commissioning Verify Operation Screen a Screen C Section Display and OperationDisplay Screen BKeypad Information Screen MessagesSecurity How the Security Code WorksThis page left blank intentionally Section Operation with Model Making Hart related settings from the keypad Off-line and On-line OperationGeneral Changing START-UP SettingsSection Programming the Transmitter Definitions Choosing and Configuring the Analytical MeasurementPurpose Choose Measurement Choose Soln Temp Corr or Sensor IsoptntlPurpose Choose Calib or Config Setting a Security CodeProcedure Setting a security code Procedure Installing default settings Making Hart Related SettingsResetting Factory Calibration and Factory Default Settings Choose ResetTransmitterChoose Default Display Selecting a Default Screen and Screen ContrastProcedure Choosing a display screen Procedure Changing screen contrastChoosing a Display Timeout Procedure Programming the display timeoutIntroduction Section Calibration TemperatureCalibrating Temperature Procedure Section Calibration pH and ORPCalibration Slope and Offset Procedure Auto Buffer Calibration Choose BufferCalProcedure Manual TWO-POINT Buffer Calibration Procedure Standardization Choose StandardizeProcedure Entering a Known Slope Value Preparation of ORP standard solutions ORP CalibrationThis page left blank intentionally Overview Section MaintenanceTransmitter Maintenance Frequency of Cleaning Cleaning ProceduresChecking the Reference Electrode Rejuvenating Reference ElectrodesGeneral ORP Sensor MaintenanceCalibration Calibration FrequencyPower Module Replacement Shipping Considerations Handling ConsiderationsEnvironmental Considerations Section Return of Material GeneralThis page left blank intentionally Warranty Specifications subject to change without notice