Emerson 6081-P instruction manual Procedure Manual TWO-POINT Buffer Calibration

Page 48

MODEL 6081-P pH/ORP

SECTION 10.0

 

CALIBRATION – PH

10.3 PROCEDURE — MANUAL TWO-POINT BUFFER CALIBRATION

1.Obtain two buffer solutions. Ideally, the buffer values should bracket the range of pH values to be measured.

2.Remove the pH sensor from the process liquid. If the process and buffer temperatures are appreciably different, place the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sensor has reached the buffer temperature. Thirty minutes is usually adequate. Make a note of the temperature.

3.Press MENU. The main menu appears. Choose Calibrate.

4.Choose pH.

5.Choose BufferCal.

6.Choose Manual.

7.Choose Buffer1.

8.Rinse the sensor with water and place it in buffer 1. Be sure the glass bulb and reference junction are com- pletely submerged. Swirl the sensor.

9.The reading in the top line is the live pH reading. Wait until the live reading is stable. Then, use the arrow keys to change the reading in the second line to the match the pH value of the buffer. The pH of buffer solutions is a function of temperature. Be sure to enter the pH of the buffer at the actual temperature of the buffer.

10.Remove the sensor from buffer 1 and rinse it with water. Place it in buffer 2. Be sure the glass bulb and the reference junction are completely submerged. Swirl the sensor. Choose Buffer2.

11.The reading in the top line is the live pH reading. Wait until the live reading is stable. Then, use the arrow keys to change the reading in the second line to the match the pH value of the buffer. The pH of buffer solutions is a function of temperature. Be sure to enter the pH of the buffer at the actual temperature of the buffer.

12.The screen at left appears momentarily.

13.If the calibration was successful, the transmitter will display the offset and slope (at 25°). The display will return to the screen in step 5.

14.If the slope is out of range (less than 45 mV/pH or greater than 60 mV/pH) or if the offset exceeds the value pro- grammed in Section 8.4, an error screen appears. The display then returns to the screen in step 6.

15.To return to the main display, press MENU then EXIT.

38

Image 48
Contents Model 6081-P Shipping considerations for wireless products Power Modules Essential InstructionsAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D List of Figures Number TitleIii This page left blank intentionally Section Description and Specifications Features and ApplicationsSpecifications General Specifications Wireless Functional SpecificationsMeasurement update rate 1/sec. to 1/10 min PH Range 0 toHart Communications Telecommunication ComplianceFCC and IC Asset Management Solutions Section Installation ConsiderationsUnpacking and Inspection PRE-INSTALLATION Setup Temperature ElementReference Electrode Impedance Preamplifier LocationSection Mounting on a Flat Surface Pipe Mounting Power Module Installation Installing the Power ModuleSection This page left blank intentionally Section Sensor Wiring Sensor WiringGeneral Information This page left blank intentionally Section Intrinsically Safe Installation This page left blank intentionallyThis page left blank intentionally Device Network Configuration Section CommissioningNetwork Communications Verify Operation Section Display and Operation DisplayScreen a Screen C Screen BKeypad Information Screen MessagesSecurity How the Security Code WorksThis page left blank intentionally Section Operation with Model Making Hart related settings from the keypad Off-line and On-line OperationChanging START-UP Settings Section Programming the TransmitterGeneral Choosing and Configuring the Analytical Measurement PurposeDefinitions Choose Measurement Choose Soln Temp Corr or Sensor IsoptntlPurpose Setting a Security Code Procedure Setting a security codeChoose Calib or Config Making Hart Related Settings Resetting Factory Calibration and Factory Default SettingsProcedure Installing default settings Choose ResetTransmitterSelecting a Default Screen and Screen Contrast Procedure Choosing a display screenChoose Default Display Procedure Changing screen contrastChoosing a Display Timeout Procedure Programming the display timeoutSection Calibration Temperature Calibrating TemperatureIntroduction Procedure Section Calibration pH and ORPCalibration Slope and Offset Procedure Auto Buffer Calibration Choose BufferCalProcedure Manual TWO-POINT Buffer Calibration Procedure Standardization Choose StandardizeProcedure Entering a Known Slope Value Preparation of ORP standard solutions ORP CalibrationThis page left blank intentionally Section Maintenance Transmitter MaintenanceOverview Frequency of Cleaning Cleaning ProceduresChecking the Reference Electrode Rejuvenating Reference ElectrodesORP Sensor Maintenance CalibrationGeneral Calibration FrequencyPower Module Replacement Handling Considerations Environmental ConsiderationsShipping Considerations Section Return of Material GeneralThis page left blank intentionally Warranty Specifications subject to change without notice