Page 37

MODEL 6081-P pH/ORP

SECTION 8.0

 

PROGRAMMING THE TRANSMITTER

8.3.3 Procedure to configure: Measurement.

To choose a menu item, move the cursor to the item and press ENTER.

To store a number or setting, press ENTER.

1.Press MENU. The main menu screen appears. Choose Program.

2.Choose Measurement.

3.Choose pH, Redox, or ORP.

If you chose pH, do steps 5 through 9. If you chose ORP or Redox, do step 10.

4.Enter the correct preamplifier location. The default setting is within the transmitter.

5.For measure sampling rate, select 1 second to 10 minutes by scrolling through choices. Press ENTER.

6.Choose Soln Temp Corr or Sensor Isoptntl.

7.For Soln Temp Corr, choose Off, UltraPure, HighpH, or Custom. For Custom, enter the desired tempera- ture coefficient.

8.For Sensor Isoptntl, enter the desired sensor isopotential pH. Do not change the sensor isopotential pH unless the sensor is known to have an isopotential pH different from 7.00.

9.Choose Low or High Reference Impedance to match the installed sensor’s reference impedance signal. The default setting is Low Impedance to match standard pH sensors. Press EXIT twice to return to the Program menu.

10.If Redox or ORP was selected, there are no further settings to make. Press EXIT to return to the Program menu.

11.To return to the main display, press MENU followed by EXIT.

27

Image 37
Contents Model 6081-P Essential Instructions Shipping considerations for wireless products Power ModulesAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D Number Title List of FiguresIii This page left blank intentionally Features and Applications Section Description and SpecificationsSpecifications General Functional Specifications Specifications WirelessMeasurement update rate 1/sec. to 1/10 min PH Range 0 toTelecommunication Compliance Hart CommunicationsFCC and IC Asset Management Solutions Considerations Section InstallationUnpacking and Inspection Temperature Element PRE-INSTALLATION SetupReference Electrode Impedance Preamplifier LocationSection Mounting on a Flat Surface Pipe Mounting Installing the Power Module Power Module InstallationSection This page left blank intentionally Sensor Wiring Section Sensor WiringGeneral Information This page left blank intentionally This page left blank intentionally Section Intrinsically Safe InstallationThis page left blank intentionally Section Commissioning Device Network ConfigurationNetwork Communications Verify Operation Display Section Display and OperationScreen a Screen C Screen BInformation Screen Messages KeypadHow the Security Code Works SecurityThis page left blank intentionally Section Operation with Model Off-line and On-line Operation Making Hart related settings from the keypadSection Programming the Transmitter Changing START-UP SettingsGeneral Purpose Choosing and Configuring the Analytical MeasurementDefinitions Choose Soln Temp Corr or Sensor Isoptntl Choose MeasurementPurpose Procedure Setting a security code Setting a Security CodeChoose Calib or Config Resetting Factory Calibration and Factory Default Settings Making Hart Related SettingsProcedure Installing default settings Choose ResetTransmitterProcedure Choosing a display screen Selecting a Default Screen and Screen ContrastChoose Default Display Procedure Changing screen contrastProcedure Programming the display timeout Choosing a Display TimeoutCalibrating Temperature Section Calibration TemperatureIntroduction Procedure Calibration pH and ORP SectionCalibration Slope and Offset Choose BufferCal Procedure Auto Buffer CalibrationProcedure Manual TWO-POINT Buffer Calibration Choose Standardize Procedure StandardizationProcedure Entering a Known Slope Value ORP Calibration Preparation of ORP standard solutionsThis page left blank intentionally Transmitter Maintenance Section MaintenanceOverview Cleaning Procedures Frequency of CleaningRejuvenating Reference Electrodes Checking the Reference ElectrodeCalibration ORP Sensor MaintenanceGeneral Calibration FrequencyPower Module Replacement Environmental Considerations Handling ConsiderationsShipping Considerations General Section Return of MaterialThis page left blank intentionally Warranty Specifications subject to change without notice