SECTION 11.0

 

MAINTENANCE

11.4 ORP SENSOR MAINTENANCE

11.4.1 Frequency of Cleaning

The frequency at which an ORP sensor should be inspected and cleaned can be determined only by experience. If the process liquid coats or fouls the sensor, frequent cleaning may be necessary. If the process does not contain a high level of suspended solids, the need for regular cleaning will be less.

11.4.2 Cleaning Procedures

The platinum electrode is easily cleaned by using a tissue to rub the metal surface with a paste of baking soda (sodium bicarbonate). A clean platinum electrode is bright and shiny.

11.4.3 Checking the Reference Electrode

ORP electrodes manufactured by Rosemount Analytical have a silver/silver chloride reference. Section 12.3.3 describes how to check the performance of the reference electrode.

11.5 CALIBRATION

11.5.1 General

Many users regard calibration as a routine part of sensor/transmitter maintenance. Procedures for calibrating pH sensors, ORP sensors, and general information regarding the use of pH calibration buffers and ORP standards are given in Sections 10.0 Calibration of pH and ORP Measurements.

11.5.2 Calibration Frequency

The frequency at which sensors should be calibrated can be determined only by experience. Many factors influ- ence calibration frequency. Sensors installed in dirty or corrosive process streams usually require more frequent calibration than sensors used in clean water. Sensors measuring extreme pH values, particularly high pH, also require more frequent calibration than sensors measuring mid-range pH. The width of the pH or ORP control range and the consequences of an out-of-limits condition has a major influence on calibration frequency. The nar- rower the control range and the greater the sensitivity of the process to control excursions, the more often the sensor should be checked. Finally, if monitoring data are reported to regulatory agencies, the agency itself may dictate the calibration frequency.

Use the following procedure to determine how often a pH sensor should be calibrated.

1.Calibrate the sensor. Record the date of calibration and the sensor response in buffers. That is, after calibrat- ing, place the sensor back in the buffers and record the pH and temperature reading in each buffer. Also note the value of the reference offset and slope.

2.Install the sensor in the process stream.

3.After the appropriate period—two weeks for a clean process, several days for a dirty or aggressive process— remove the sensor and check its performance in buffers. Record the pH and temperature readings. The per- formance of the sensor in buffer after it has been in service is called the as-found condition. Keeping a good record of as-found data is an important step in determining the calibration frequency.

4.If the as-found data are acceptable, do not recalibrate the sensor. Return it to the process. Continue checking the calibration at the same interval.

5.If the as-found data are not acceptable, recalibrate the sensor. After calibration, check the sensor response in each buffer and record the results. Also note the reference offset and the slope. Return the sensor to service. Check the sensor again after a period shorter than the one originally selected. For example, if the first interval was two weeks, repeat the check after one week.

6.After a while it will become apparent how long the sensor holds calibration. The minimum calibration frequency can then be determined.

7.Check the calibration of the sensor at least several times during the regular calibration interval. Interim checks verify the sensor is still in calibration and validate the process measurements made since the last calibration or calibration check.

46

Image 56
Contents Model 6081-P Shipping considerations for wireless products Power Modules Essential InstructionsAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D Iii List of FiguresNumber Title This page left blank intentionally Specifications General Section Description and SpecificationsFeatures and Applications Specifications Wireless Functional SpecificationsMeasurement update rate 1/sec. to 1/10 min PH Range 0 toFCC and IC Hart CommunicationsTelecommunication Compliance Asset Management Solutions Unpacking and Inspection Section InstallationConsiderations PRE-INSTALLATION Setup Temperature ElementReference Electrode Impedance Preamplifier LocationSection Mounting on a Flat Surface Pipe Mounting Power Module Installation Installing the Power ModuleSection This page left blank intentionally General Information Section Sensor WiringSensor Wiring This page left blank intentionally Section Intrinsically Safe Installation This page left blank intentionallyThis page left blank intentionally Network Communications Device Network ConfigurationSection Commissioning Verify Operation Section Display and Operation DisplayScreen a Screen C Screen BKeypad Information Screen MessagesSecurity How the Security Code WorksThis page left blank intentionally Section Operation with Model Making Hart related settings from the keypad Off-line and On-line OperationGeneral Changing START-UP SettingsSection Programming the Transmitter Definitions Choosing and Configuring the Analytical MeasurementPurpose Choose Measurement Choose Soln Temp Corr or Sensor IsoptntlPurpose Choose Calib or Config Setting a Security CodeProcedure Setting a security code Making Hart Related Settings Resetting Factory Calibration and Factory Default SettingsProcedure Installing default settings Choose ResetTransmitterSelecting a Default Screen and Screen Contrast Procedure Choosing a display screenChoose Default Display Procedure Changing screen contrastChoosing a Display Timeout Procedure Programming the display timeoutIntroduction Section Calibration TemperatureCalibrating Temperature Procedure Section Calibration pH and ORPCalibration Slope and Offset Procedure Auto Buffer Calibration Choose BufferCalProcedure Manual TWO-POINT Buffer Calibration Procedure Standardization Choose StandardizeProcedure Entering a Known Slope Value Preparation of ORP standard solutions ORP CalibrationThis page left blank intentionally Overview Section MaintenanceTransmitter Maintenance Frequency of Cleaning Cleaning ProceduresChecking the Reference Electrode Rejuvenating Reference ElectrodesORP Sensor Maintenance CalibrationGeneral Calibration FrequencyPower Module Replacement Shipping Considerations Handling ConsiderationsEnvironmental Considerations Section Return of Material GeneralThis page left blank intentionally Warranty Specifications subject to change without notice