Emerson 6081-P instruction manual Procedure Auto Buffer Calibration, Choose BufferCal

Page 47

MODEL 6081-P pH/ORP

SECTION 10.0

 

CALIBRATION – PH

10.2 PROCEDURE — AUTO BUFFER CALIBRATION

1.Obtain two buffer solutions. Ideally, the buffer values should bracket the range of pH values to be measured.

2.Remove the pH sensor from the process liquid. If the process and buffer temperatures are appreciably differ- ent, place the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sensor has reached the buffer temperature. Thirty minutes is usually adequate.

3.Press MENU. The main menu appears. Choose Calibrate.

4.Choose pH.

5.Choose BufferCal.

6.Choose Auto.

7.To continue with the calibration, choose Buffer1.Then go to step 8. To change stability criteria, choose Setup and go to step 19.

8.Rinse the sensor with water and place it in buffer 1. Be sure the glass bulb and the reference junction are com- pletely submerged. Swirl the sensor.

9.The screen at left is displayed with “Wait” flashing until the reading is stable. The default stability setting is <0.02 pH change in 10 sec. To change the stability criteria, go to step 19. When the reading is stable, the screen in step 10 appears.

10.The top line shows the actual reading. The transmitter also identifies the buffer and displays the nominal buffer value (buffer pH at 25°C). If the displayed value is not correct, press ￿ or ￿ to display the correct value. The nominal value will change, for example from 7.01 to 6.86 pH. Press ENTER to store.

11.The screen at left appears momentarily.

12.The screen at left appears. Remove the sensor from Buffer 1, rinse it with water, and place it in Buffer 2. Be sure the glass bulb and the reference junction are completely submerged. Swirl the sensor. Choose Buffer2.

13.The screen at left is displayed with “Wait” flashing until the reading is stable. When the reading is stable, the screen in step 14 appears.

14.The top line shows the actual reading. The transmitter also identifies the buffer and displays the nominal buffer value (buffer pH at 25°C). If the displayed value is not correct, press ￿ or ￿ to display the correct value. The nominal value will change, for example from 9.91 to 10.02 pH. Press ENTER to store.

15.The screen at the left appears momentarily.

16.If the calibration was successful, the transmitter will display the offset and slope (at 25°). The display will return to the screen in step 6.

17.If the slope is out of range (less than 45 mV/pH or greater than 60 mV/pH) or if the offset exceeds the value pro- grammed in Section 8.4, an error screen appears. The display then returns to the screen in step 6.

18.To return to the main display, press MENU then EXIT.

19.Choosing Setup in step 7 causes the Buffer Stabilize screen to appear. The transmitter will not accept cali- bration data until the pH reading is stable. The default requirement is a pH change less than 0.02 units in 10 seconds. To change the stability criteria:

a.Enter the desired stabilization time

b.Enter the minimum amount the reading is permitted to change in the time specified in step 19a.

20.To return to the main display, press MENU then EXIT.

37

Image 47
Contents Model 6081-P Essential Instructions Shipping considerations for wireless products Power ModulesAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D Iii List of FiguresNumber Title This page left blank intentionally Specifications General Section Description and SpecificationsFeatures and Applications PH Range 0 to Specifications WirelessFunctional Specifications Measurement update rate 1/sec. to 1/10 minFCC and IC Hart CommunicationsTelecommunication Compliance Asset Management Solutions Unpacking and Inspection Section InstallationConsiderations Preamplifier Location PRE-INSTALLATION SetupTemperature Element Reference Electrode ImpedanceSection Mounting on a Flat Surface Pipe Mounting Installing the Power Module Power Module InstallationSection This page left blank intentionally General Information Section Sensor WiringSensor Wiring This page left blank intentionally This page left blank intentionally Section Intrinsically Safe InstallationThis page left blank intentionally Network Communications Device Network ConfigurationSection Commissioning Verify Operation Screen B Section Display and OperationDisplay Screen a Screen CInformation Screen Messages KeypadHow the Security Code Works SecurityThis page left blank intentionally Section Operation with Model Off-line and On-line Operation Making Hart related settings from the keypadGeneral Changing START-UP SettingsSection Programming the Transmitter Definitions Choosing and Configuring the Analytical MeasurementPurpose Choose Soln Temp Corr or Sensor Isoptntl Choose MeasurementPurpose Choose Calib or Config Setting a Security CodeProcedure Setting a security code Choose ResetTransmitter Making Hart Related SettingsResetting Factory Calibration and Factory Default Settings Procedure Installing default settingsProcedure Changing screen contrast Selecting a Default Screen and Screen ContrastProcedure Choosing a display screen Choose Default DisplayProcedure Programming the display timeout Choosing a Display TimeoutIntroduction Section Calibration TemperatureCalibrating Temperature Procedure Calibration pH and ORP SectionCalibration Slope and Offset Choose BufferCal Procedure Auto Buffer CalibrationProcedure Manual TWO-POINT Buffer Calibration Choose Standardize Procedure StandardizationProcedure Entering a Known Slope Value ORP Calibration Preparation of ORP standard solutionsThis page left blank intentionally Overview Section MaintenanceTransmitter Maintenance Cleaning Procedures Frequency of CleaningRejuvenating Reference Electrodes Checking the Reference ElectrodeCalibration Frequency ORP Sensor MaintenanceCalibration GeneralPower Module Replacement Shipping Considerations Handling ConsiderationsEnvironmental Considerations General Section Return of MaterialThis page left blank intentionally Warranty Specifications subject to change without notice
Related manuals
Manual 16 pages 28.1 Kb