Emerson 6081-P instruction manual ORP Calibration, Preparation of ORP standard solutions

Page 51

MODEL 6081-P pH/ORP

SECTION 10.0

 

CALIBRATION – PH

10.6 ORP CALIBRATION

10.6.1 Purpose

1.For process control, it is often important to make the measured ORP agree with the ORP of a standard solu- tion.

2.During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point.

10.6.2 Preparation of ORP standard solutions

ASTM D1498-93 gives procedures for the preparation of iron (II) - iron (III) and quinhydrone ORP standards. The iron (II) - iron (III) standard is recommended. It is fairly easy to make, is not particularly hazardous, and has a shelf life of about one year. In contrast, quinhydrone standards contain toxic quinhydrone and have only an eight-hour shelf life.

Iron (II) - iron (III) standard is available from Rosemount Analytical as PN R508-16OZ. The ORP of the standard solution measured against a silver-silver chloride reference electrode is 476±20mV at 25°C. The redox potential is -476±20mV at 25°C.

10.6.3 Procedure

1.Press MENU. The main menu screen appears. Choose Calibrate.

2.Choose ORP.

3.The top line shows the actual ORP or redox potential (Live). Once the reading is stable, change the number in the second line to the desired value. Press ENTER.

4.The screen on the left will appear briefly.

5.The display returns to the Cal Sensor screen. Press EXIT. Choose the other sensor and repeat steps 2 through 4.

41

Image 51
Contents Model 6081-P Essential Instructions Shipping considerations for wireless products Power ModulesAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D List of Figures Number TitleIii This page left blank intentionally Section Description and Specifications Features and ApplicationsSpecifications General PH Range 0 to Specifications WirelessFunctional Specifications Measurement update rate 1/sec. to 1/10 minHart Communications Telecommunication ComplianceFCC and IC Asset Management Solutions Section Installation ConsiderationsUnpacking and Inspection Preamplifier Location PRE-INSTALLATION SetupTemperature Element Reference Electrode ImpedanceSection Mounting on a Flat Surface Pipe Mounting Installing the Power Module Power Module InstallationSection This page left blank intentionally Section Sensor Wiring Sensor WiringGeneral Information This page left blank intentionally This page left blank intentionally Section Intrinsically Safe InstallationThis page left blank intentionally Device Network Configuration Section CommissioningNetwork Communications Verify Operation Screen B Section Display and OperationDisplay Screen a Screen CInformation Screen Messages KeypadHow the Security Code Works SecurityThis page left blank intentionally Section Operation with Model Off-line and On-line Operation Making Hart related settings from the keypadChanging START-UP Settings Section Programming the TransmitterGeneral Choosing and Configuring the Analytical Measurement PurposeDefinitions Choose Soln Temp Corr or Sensor Isoptntl Choose MeasurementPurpose Setting a Security Code Procedure Setting a security codeChoose Calib or Config Choose ResetTransmitter Making Hart Related SettingsResetting Factory Calibration and Factory Default Settings Procedure Installing default settingsProcedure Changing screen contrast Selecting a Default Screen and Screen ContrastProcedure Choosing a display screen Choose Default DisplayProcedure Programming the display timeout Choosing a Display TimeoutSection Calibration Temperature Calibrating TemperatureIntroduction Procedure Calibration pH and ORP SectionCalibration Slope and Offset Choose BufferCal Procedure Auto Buffer CalibrationProcedure Manual TWO-POINT Buffer Calibration Choose Standardize Procedure StandardizationProcedure Entering a Known Slope Value ORP Calibration Preparation of ORP standard solutionsThis page left blank intentionally Section Maintenance Transmitter MaintenanceOverview Cleaning Procedures Frequency of CleaningRejuvenating Reference Electrodes Checking the Reference ElectrodeCalibration Frequency ORP Sensor MaintenanceCalibration GeneralPower Module Replacement Handling Considerations Environmental ConsiderationsShipping Considerations General Section Return of MaterialThis page left blank intentionally Warranty Specifications subject to change without notice