CALIBRATION – TEMPERTURE

9.2.2 Procedure

1.Remove the sensor from the process. Place it in an insulated container of water along with a calibrated ther- mometer. Submerge at least the bottom two inches of the sensor. Stir continuously.

2.Allow the sensor to reach thermal equilibrium. For some sensors, the time constant for a change in tempera- ture is 5 min., so it may take as long as 30 min. for temperature equilibration.

3.If the sensor cannot be removed from the process, measure the temperature of a flowing sample taken from a point as close to the sensor as possible. Let the sample continuously overflow an insulated container hold- ing a calibrated thermometer.

4.Change the 6081 display to match the calibrated thermometer using the procedure below.

1.Press MENU. The menu screen appears. Choose Calibrate.

2.Choose Temp.

3.If transmitter was programmed in Section 8.4 to use the actual process temperature, go to step 7. If the transmitter was programmed to use a temperature entered by the user, go to step 9.

4.To calibrate the temperature, change the number in the second line to match the temperature measured with the standard thermometer. Press ENTER.

5.Press MENU then EXIT to return to the main display.

6.If the temperature value shown in the display is not correct, use the arrow keys to change it to the desired value. The transmitter will use the temperature entered in this step in all measurements and calculations, no matter what the true temperature is.

7.Press MENU then EXIT to return to the main display.

34

Image 44
Contents Model 6081-P Shipping considerations for wireless products Power Modules Essential InstructionsAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D Iii List of FiguresNumber Title This page left blank intentionally Specifications General Section Description and SpecificationsFeatures and Applications Specifications Wireless Functional SpecificationsMeasurement update rate 1/sec. to 1/10 min PH Range 0 toFCC and IC Hart CommunicationsTelecommunication Compliance Asset Management Solutions Unpacking and Inspection Section InstallationConsiderations PRE-INSTALLATION Setup Temperature ElementReference Electrode Impedance Preamplifier LocationSection Mounting on a Flat Surface Pipe Mounting Power Module Installation Installing the Power ModuleSection This page left blank intentionally General Information Section Sensor WiringSensor Wiring This page left blank intentionally Section Intrinsically Safe Installation This page left blank intentionallyThis page left blank intentionally Network Communications Device Network ConfigurationSection Commissioning Verify Operation Section Display and Operation DisplayScreen a Screen C Screen BKeypad Information Screen MessagesSecurity How the Security Code WorksThis page left blank intentionally Section Operation with Model Making Hart related settings from the keypad Off-line and On-line OperationGeneral Changing START-UP SettingsSection Programming the Transmitter Definitions Choosing and Configuring the Analytical MeasurementPurpose Choose Measurement Choose Soln Temp Corr or Sensor IsoptntlPurpose Choose Calib or Config Setting a Security CodeProcedure Setting a security code Making Hart Related Settings Resetting Factory Calibration and Factory Default SettingsProcedure Installing default settings Choose ResetTransmitterSelecting a Default Screen and Screen Contrast Procedure Choosing a display screenChoose Default Display Procedure Changing screen contrastChoosing a Display Timeout Procedure Programming the display timeoutIntroduction Section Calibration TemperatureCalibrating Temperature Procedure Section Calibration pH and ORPCalibration Slope and Offset Procedure Auto Buffer Calibration Choose BufferCalProcedure Manual TWO-POINT Buffer Calibration Procedure Standardization Choose StandardizeProcedure Entering a Known Slope Value Preparation of ORP standard solutions ORP CalibrationThis page left blank intentionally Overview Section MaintenanceTransmitter Maintenance Frequency of Cleaning Cleaning ProceduresChecking the Reference Electrode Rejuvenating Reference ElectrodesORP Sensor Maintenance CalibrationGeneral Calibration FrequencyPower Module Replacement Shipping Considerations Handling ConsiderationsEnvironmental Considerations Section Return of Material GeneralThis page left blank intentionally Warranty Specifications subject to change without notice