Emerson 6081-P instruction manual Frequency of Cleaning, Cleaning Procedures

Page 54

MODEL 6081-P pH/ORP

SECTION 11.0

 

MAINTENANCE

11.3 pH SENSOR MAINTENANCE

11.3.1 Frequency of Cleaning

The frequency at which a sensor should be inspected and cleaned can be determined only by experience. If the process liquid coats or fouls the sensor, frequent cleaning may be necessary. If the process does not contain a high level of suspended solids, the need for regular cleaning will be less. Often an increase in glass impedance indicates the electrode is becoming fouled and needs cleaning. Refer to Section 12.4 for a description of the glass impedance diagnostic.

11.3.2 Cleaning Procedures

PROBLEM

CLEANING SUGGESTIONS

 

 

Loose scale or debris

Use a stream of water from a wash bottle to rinse away solids from the tip of the

 

 

sensor. If water does not work, gently wipe the glass bulb and liquid junction with a

 

soft cloth, tissue, cotton-tipped swab, or a soft bristle brush.

 

 

Oil and grease

Wash the glass bulb with mild detergent solution and rinse thoroughly with water.

 

 

Hard scale (carbonate sulfate

If wiping the sensor tip with a tissue or cotton swab does not remove the scale,

 

scales and corrosion products)

soak the glass bulb ONLY in a solution of 5% hydrochloric acid. To prepare the

 

acid solution, add 15 mL of concentrated hydrochloric acid to 85 mL of water.

 

Keep the acid away from the liquid junction and from any stainless steel portions

 

of the sensor. Rinse the sensor thoroughly with deionized water. Some scales (for

 

example, calcium sulfate) cannot be removed easily with acid. Soaking the glass

 

bulb in a 2% solution of disodium EDTA may be helpful.

 

 

When using acid or alkaline solvents, be careful to keep the solvent away from the liquid junction. If the cleaning sol- vent contacts the junction, hydrogen ions (acid solvent) or hydroxide ions (alkaline solvent) will diffuse into the junc- tion. Because hydrogen and hydroxide ions have much greater mobility than other ions, they produce a large junction potential. When the electrode goes back in service, the hydrogen or hydroxide ions slowly diffuse out of the junction, causing the liquid junction potential and the pH reading to drift. It may take hours or days for the reading to stabilize. For a discussion of the influence of ion mobility on liquid junction potentials, see Section 13.4.

Consult the sensor instruction manual for additional information.

Always recalibrate the sensor after cleaning. If the sensor was cleaned with detergent or acid, soak the sensor in pH 4 or pH 7 buffer for at least an hour before calibrating.

44

Image 54
Contents Model 6081-P Shipping considerations for wireless products Power Modules Essential InstructionsAbout This Document Quick Start Guide Hart PollAddress Menu Tree for Model 6081 pH Wireless Transmitter Model 6081-P pH/ORP Table of Contents CONT’D List of Figures Number TitleIii This page left blank intentionally Section Description and Specifications Features and ApplicationsSpecifications General Measurement update rate 1/sec. to 1/10 min Specifications WirelessFunctional Specifications PH Range 0 toHart Communications Telecommunication ComplianceFCC and IC Asset Management Solutions Section Installation ConsiderationsUnpacking and Inspection Reference Electrode Impedance PRE-INSTALLATION SetupTemperature Element Preamplifier LocationSection Mounting on a Flat Surface Pipe Mounting Power Module Installation Installing the Power ModuleSection This page left blank intentionally Section Sensor Wiring Sensor WiringGeneral Information This page left blank intentionally Section Intrinsically Safe Installation This page left blank intentionallyThis page left blank intentionally Device Network Configuration Section CommissioningNetwork Communications Verify Operation Screen a Screen C Section Display and OperationDisplay Screen BKeypad Information Screen MessagesSecurity How the Security Code WorksThis page left blank intentionally Section Operation with Model Making Hart related settings from the keypad Off-line and On-line OperationChanging START-UP Settings Section Programming the TransmitterGeneral Choosing and Configuring the Analytical Measurement PurposeDefinitions Choose Measurement Choose Soln Temp Corr or Sensor IsoptntlPurpose Setting a Security Code Procedure Setting a security codeChoose Calib or Config Procedure Installing default settings Making Hart Related SettingsResetting Factory Calibration and Factory Default Settings Choose ResetTransmitterChoose Default Display Selecting a Default Screen and Screen ContrastProcedure Choosing a display screen Procedure Changing screen contrastChoosing a Display Timeout Procedure Programming the display timeoutSection Calibration Temperature Calibrating TemperatureIntroduction Procedure Section Calibration pH and ORPCalibration Slope and Offset Procedure Auto Buffer Calibration Choose BufferCalProcedure Manual TWO-POINT Buffer Calibration Procedure Standardization Choose StandardizeProcedure Entering a Known Slope Value Preparation of ORP standard solutions ORP CalibrationThis page left blank intentionally Section Maintenance Transmitter MaintenanceOverview Frequency of Cleaning Cleaning ProceduresChecking the Reference Electrode Rejuvenating Reference ElectrodesGeneral ORP Sensor MaintenanceCalibration Calibration FrequencyPower Module Replacement Handling Considerations Environmental ConsiderationsShipping Considerations Section Return of Material GeneralThis page left blank intentionally Warranty Specifications subject to change without notice